![GSi TYMPSTAR PRO Скачать руководство пользователя страница 92](http://html1.mh-extra.com/html/gsi/tympstar-pro/tympstar-pro_user-manual_2275120092.webp)
automatically identified by the cursor on the graphic displays where
Δ
B
crosses 0 mmhos. Resonance of the ear is reached when stiffness and
mass
components are equal in magnitude
.
The Delta Peak-Start table contains the numerical values for the graphical results
obtained at resonant frequency. The table includes the resonant frequency (Hz),
Δ
B (mmho) and
Ø
B (degrees). The starting pressure and peak in daPa are
also displayed.
Tab 3-Tympanogram at Resonant Frequency
The tympanogram obtained at resonant frequency is displayed in the graph
(immittance component B). The table on lower right side of the graph displays
the results of the measurement. The header of the table indicates the resonant
frequency. The Tymp column displays the direction of the pressure sweep and
the immittance component that is being shown on the tympanogram. The Vea
column displays the acoustic equivalent volume (in cm
3
). The Peak column
displays the pressure (daPa) and admittance (mmho) value of the peak of the
tympanogram.