around the curve. If no threshold is found, the test will be marked as “no
response” (NR). When Threshold Seek is not enabled, the data may be stored by
the user by selecting the Store icon or the NR-Store icon. If Store is selected and
a threshold is selected in the scratchpad, the threshold will be indicated in the
stored reflex. If no threshold is selected in the scratchpad, the test will be “not
set” (NS). If NR-Store is selected, the stored reflex will be marked as no
response (NR). The reflex may also be stored automatically by the system. This
occurs when threshold seek is not enabled and the stimulus is changed or the test
is stopped. In this case, if a threshold is selected in the scratchpad, it will be
marked in the saved reflex results. If no threshold is marked, the test will be “not
set” (NS). “Not set” indicates no threshold was selected and the test has not
been marked as no response. For tests that are “not set,” selecting the threshold
or marking as a no response is required to indicate the result of the reflex.
Probe Status Indicator
The probe status indicator indicates the state of the probe and displays messages
regarding the current test. On the left side of the indicator, the two circles will be
colored to match the LED lights on the probe box. The probe may be in the
following states:
Indicator Lights
Status
Blinking Green
Ready or Test Started waiting for Present
Solid Green
Test in Progress
Blinking Green
Hold – ready to continue test
Solid Amber
Occlusion
Blinking Amber
Leak (or possible internal error on the device)
Admittance and Pressure Meters
The admittance meter displays the real time admittance as it is being recorded
from the probe in the ear canal. The units are millimhos (mmho). The current
value is displayed at the top left of the meter.