Appendix 3: Error Messages
Error - Invalid Start Direction
When recording tympanograms, the pressure sweep direction is determined by
the direction of the START button that is pressed. The direction is from a
positive pressure to a negative pressure (Left Arrow START) or from a negative
pressure to a positive pressure (Right Arrow START). The starting pressure is
determined from the configuration for the initial tympanogram or from the
ending hold pressure if multiple tympanograms are being collected. If the
starting or hold pressure is positive then the instrument is not able to pressurize in
a positive direction and therefore if you press the Right Arrow START button the
error message is displayed. Likewise, if the starting or hold pressure is negative
then the instrument is not able to pressurize in a negative direction and therefore
if you press the Left Arrow START button the error message is displayed
Error - Please select a Curve or Curves from the Scratchpad to Store
After an acoustic reflex has been recorded the data is displayed as a curve or
curves in the scratchpad. In order to store a reflex as a threshold, the curve in the
scratch pad is selected then stored as a threshold. To select a curve on the
scratchpad, the curve is touched on the LCD or clicked with the mouse. If there
is no curve selected and the Store menu item is pressed the message is displayed.
Error – No Data to Print
Data collected from the tests is available to print. If there is no data or a non data
screen is displayed such as the configuration screen you may receive the “No
data to Print” error message.
XXX is not available:
Indicates a front panel button was pressed which is
currently not available. This typically is due to a test currently in process.
Printer communication error:
If communications problems occur during the
course of printing, an error message will be displayed.
Calibration Error
: Both the Probe (IPSI) and Contra transducers are required to
be calibrated. If this error appears, it is recommended to contact your GSI
Service Representative.
Multi-Hz Calibration Error -- The system has not been calibrated for Multi-Hz
testing. Contact your GSI service representative to calibrate.
Altitude Change Error -- Changing the altitude requires a new Multi-Hz
calibration at the new altitude. Contact your GSI service representative to
calibrate.
Error:
If there are general system errors, a dialog box with “Error” in the title
will be shown with the given error.