Tympanometry Test Results
The lower right side of the Tympanometry screen displays the tympanogram
results in a table. The header of the table indicates the test number and the probe
frequency. The Tymp column displays the tympanogram number (up to 3 per
test), the direction of the pressure sweep and the component that is being shown
on the tympanogram. The Vea column displays the acoustic equivalent volume
(in cm
3
). The Peak column displays the pressure (daPa) and admittance (mmho)
value of the indicated peak of the tympanogram. The Tymp Measure t column
displays the tympanometric width (daPa) or gradient (mmho). The gradient
option is selected from the Settings dialog and is only displayed for the 226 Hz
probe tone and Y immittance component. The Type column displays
tympanogram classification type that the user selects from the Set Type link in
the column.
Set Type
The Set Type option displays a dialog with the tympanogram types that are
available to be selected. Separate tympanogram types may be defined for Y and
B/G components and are defined in the PC configuration software (TympStar Pro
Config App).