As an alternative to disinfecting wipes, an ultraviolet (UV) light source could be
used to disinfect touch screen surfaces.
Cleaning patient contact devices
Patient contact parts should be cleaned prior to each use. The probe and
contralateral earphone and probe box may be wiped with a slightly damp cloth
containing soap and water ammonia based cleaners or bleach based cleaners.
Gently wipe the items with the slightly damp cloth taking care not to get moisture
in the speaker portion of the probe and contralateral earphones. The eartips for
the probe are for single use only and should not be reused.
Probe care
With normal use, cerumen can work its way inside the probe nose cone (probe
tip). During the warm-up period each day and throughout the day, inspect the
probe tip to make sure it is clean and free of cerumen. Refer to the following
instructions for cleaning and maintaining the instrument’s probe.
1:
Unscrew the probe cap.
Содержание TYMPSTAR PRO