Series 400 / 402 Gas Chromatograph Operating Manual GOW-MAC Instrument Co.
May 2017, Rev. 7
Chapter 9 - Maintenance
9.1 The Detector and Elements
A. Limits of Operation
The elements (aka hotwires or fi laments) used in the Series 400 GC are GOW-MAC
rhenium-tungsten (WX). They may be purchased from GOW-MAC in "quads". A quad
consists of four matched and balanced fi laments with cable (P/N 13-355-400). WX
fi laments should be used within the specifi ed limits as shown in Chart 4-1.
B. Cleaning the Cell
Imperfectly vaporized liquid samples may be deposited on the fi laments or cavity walls of
the TC cell. Such deposits will aff ect the heat transfer properties of the cell and change the
zero balance progressively until the zero will no longer be suffi cient. Contamination of the
detector may be removed by the following method
AFTER THE OVEN HAS COOLED
COMPLETELY:
1. Shut "OFF" carrier gas and electrical power to the GC.
2. Remove columns from the instrument.
3. Replace columns with lengths of metal tubing (Figure 9-1).
4. Inject 100 mL of acetone through the septum and leave overnight.
WARNING
!
CLEANING SOLVENTS ARE EXTREMELY FLAMMABLE.
USE CARE WHEN USING THESE MATERIALS. DO NOT
EXPOSE THEM TO OPEN FLAMES OR SMOKING
MATERIALS. DISPOSE OF SOLVENTS PROPERLY.
Empty metal
columns
Figure 9-1
Содержание 400 Series
Страница 2: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 2 ...
Страница 4: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 4 ...
Страница 8: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 8 ...
Страница 34: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 34 ...
Страница 40: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 40 ...
Страница 56: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 56 ...
Страница 58: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 58 ...
Страница 60: ...GOW MAC Instrument Co 60 ...
Страница 62: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 62 ...
Страница 64: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 64 ...