![Gow-Mac 400 Series Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/gow-mac/400-series/400-series_operating-manual_2248315037.webp)
37
Series 400 / 402 Gas Chromatograph Operating Manual GOW-MAC Instrument Co.
May 2017, Rev. 7
B. Procedure for Removing and Replacing Loops
1. Unscrew loop mounting fi ttings (2) and remove loop.
2. Insert new loop and tighten fi ttings.
6.6 Pneumatically Actuated Valves
ACT. AIR IN
SAMPLE OUT
SAMPLE IN
VENT
VENT
SAMPLE OUT
SAMPLE IN
ACT. AIR IN
MIN. PRESSURE 15 PSIG
INLET
Sample Gas regulated to 20 psig.
Set flow to about 100 ccpm using
customer supplied needle valve.
Two micron (2
μ
) filter.
Valve Actuator Gas
Compressed Air regulated
to 30-50 psig.
FREE OF OIL AND WATER!
Carrier
Gas
Sample
Gas
Carrier Gas regulated to
20 to 75 psig depending
on column length.
Two micron (2
μ
) filter.
Air
Figure 6-1: Typical Setup. Your's may vary. Refer to enclosed fl ow diagram.
A. Pneumatically actuated gas sample valves require compressed air gas that is
free of oil
and water
to function. An air pressure of 30-50 psi is recommended to fully drive the
valves.
B. A TTL Interface PCB enables the operator to control the pneumatic actuated valves that
are installed in the GC via an external controller, i.e., computing software or computing
integrator.
This external controller must be able to supply an active high (+5 V) TTL signal to the
corresponding Valve Actuator (TTL) Cable Connection located on the GC's Valve
Housing. This +5 V signal must be kept “ON” for the entire period of time the user wants to
keep the valve(s) actuated.
C. Connect the Interface Cable provided to the VALVE ACTUATORS Terminal Strip located
on the back of the
Valve Housing(s)
. Moving top to bottom along the strip, attach the color
coded wires in the following sequence:
Содержание 400 Series
Страница 2: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 2 ...
Страница 4: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 4 ...
Страница 8: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 8 ...
Страница 34: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 34 ...
Страница 40: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 40 ...
Страница 56: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 56 ...
Страница 58: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 58 ...
Страница 60: ...GOW MAC Instrument Co 60 ...
Страница 62: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 62 ...