Series 400 / 402 Gas Chromatograph Operating Manual GOW-MAC Instrument Co.
May 2017, Rev. 7
Chapter 2 - Specifi cations
The Series 400/402 Gas Chromatograph is a rugged, compact, thermal conductivity GC designed
for high capability performance while withstanding rough, student or industrial use.
The fl ow system consists of two needle valves for accurate column fl ow control, dual injection ports,
dual columns and exits.
Temperature of injection ports, column oven and detector oven can be individually controlled. The
detector is housed in an easily accessible oven assembly. The thermal conductivity detector (TCD)
contains four 32 ohm, Rhenium-tungsten (WX) hotwire fi laments (standard).
The Series 400-P/402-P GC is equipped with a microscale preparative accessory system. This GC
is equipped with a microscale prep detector with outlet temperature control. The system is based on
the description given by Mayo, Pike, et. al., in
Microscale Organic Laboratory
, E. Wiley & Son,
N. Y., 1986.
Dimensions
Overall
16.50" H (410 mm) x 12.00" W (300 mm) x 11.00" D (270 mm)
Weight
27 lbs. (12.26 kg)
Power Requirements
Series 400: 110-120 VAC, 50/60 Hz: 600 W (max), 10 A circuit breaker
Series 402: 220-240 VAC, 50/60 Hz, 600 W (max), 4 A circuit breaker
Column Oven
Dimensions:
16.19 cm H x 25.40 cm W x 18.41 cm D
Volume:
7570.67 cm
3
V
Temperature Range:
Ambient to 400 °C
Temperature Control:
solid state, time proportioning, RTD sensors, direct reading, ambient
to 400 °C
Temperature Readout: 3-1/2 digit LCD digital meter
Protection Circuit:
shuts the column oven off if the temperature rises to 30 °C over set
point
Detector Outlet
[optional with preparative detector (Series 400-P/402-P)]
Temperature Range:
Ambient to 350 °C
Temperature Control:
phase control, RTD sensor, ambient to 350 °C
Temperature Readout:
3-1/2 digit LCD digital meter
Operating Temperature: 20 °C above column temperature
Oven Fittings
Accommodates 1/8" or 1/4" o.d. metal or glass columns
Oven Capacity
Two 1/4" o.d. x 12' columns or correspondingly longer lengths of 1/8" o.d.
Содержание 400 Series
Страница 2: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 2 ...
Страница 4: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 4 ...
Страница 8: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 8 ...
Страница 34: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 34 ...
Страница 40: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 40 ...
Страница 56: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 56 ...
Страница 58: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 58 ...
Страница 60: ...GOW MAC Instrument Co 60 ...
Страница 62: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 62 ...
Страница 64: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 64 ...