![Gow-Mac 400 Series Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/gow-mac/400-series/400-series_operating-manual_2248315015.webp)
15
Series 400 / 402 Gas Chromatograph Operating Manual GOW-MAC Instrument Co.
May 2017, Rev. 7
Chapter 3 - Installation
3.1 General
The customer should read and become familiar with this section
before
proceeding with
installation.
3.2 Additional Equipment Required
A. Carrier gas cylinder with standard pressure regulator, terminating in a 1/4" or 1/8" female
NPT fi tting.
B. Recording Device: a) Potentiometric recorder with 1 mV span, 1 second response; or b) a
computing integrator; or c) proper computer software.
C. Stopwatch and a 10 mL soap bubble fl owmeter (GM Part No. 59-230) for gas fl ow
measurements. A "rotameter" may be used for direct reading if it is properly calibrated
using a soap bubble fl owmeter. Also available is a hand-held digital bubble fl owmeter,
GOW-MAC Part No. 180-567 for automatic fl ow measurement.
D. AC power source: Series 400: 600 W at 110 - 130 V, 50/60 Hz
Series 402: 600 W at 220 - 240 V, 50/60 Hz
CAUTION
Operating instructions for bother series are the same, except
for line voltage requirements. To prevent damage to the
instrument, make sure that the AC electrical outlet is the
correct voltage for your instrument BEFORE plugging it into
the outlet.
E. GOW-MAC Installation Kit (Part No. 59-400) or 1/8" o.d. clean copper tubing and
Swagelok
®
fi ttings.
F. Columns suitable to your application. The Series 400 GC comes with one (1) 20%
Carbowax 20M column and one (1) 20% DC 200 column.
3.3 Unpacking-Inspection
A. When unpacking the instrument, check it carefully for evidence of shipping damage or
rough handling. Check to ensure that all components ordered have either been supplied
or back-ordered. Notify the Company of any discrepancies. The packing box should
be retained for use if the instrument needs to be returned to the factory for repair or
modifi cation. GOW-MAC does not supply fi eld service. All repairs are made at our
Bethlehem, PA facility or by an authorized GOW-MAC representative.
Содержание 400 Series
Страница 2: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 2 ...
Страница 4: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 4 ...
Страница 8: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 8 ...
Страница 34: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 34 ...
Страница 40: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 40 ...
Страница 56: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 56 ...
Страница 58: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 58 ...
Страница 60: ...GOW MAC Instrument Co 60 ...
Страница 62: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 62 ...
Страница 64: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 64 ...