Series 400 / 402 Gas Chromatograph Operating Manual GOW-MAC Instrument Co.
May 2017, Rev. 7
!
IMPORTANT
IT IS MOST IMPORTANT THAT THE ADAPTER BE THREADED
ONTO THE EXIT PORT COMPLETELY. THE END OF THE EXIT
PORT MUST EXTEND WELL INTO THE GROUND STANDARD
TAPERED PART OF THE ADAPTER WHEN IT IS ATTACHED
TO THE EXIT PORT. THIS ARRANGEMENT WILL ALLOW
THE END OF THE EXIT PORT TO BUTT SNUGLY AGAINST
THE COLLECTION TUBE WHEN FULLY ASSEMBLED. IT IS
ABSOLUTELY ESSENTIAL THAT THIS CONTACT IS MADE.
2. When the recorder indicates that the fraction to be collected is approaching the end of
the column, the collection tube (previously cleaned and oven dried) is seated in to the
adapter.
3. Place a collection tube into the adapter by rotating the collection tube while inserting it
into the adapter. If too much pressure is applied the collection tube will unseat. It is
important to practice rotating the collection tube
GENTLY
to obtain just enough
pressure to make a good seal, but not enough to unseat the joint.
4. The preparative GC fraction is now collected as it passes through the exit port into the
collection tube. Effi cient collection is facilitated (diminished aerosol formation) by the
large heat sink aff orded by the mass of the stainless steel adapter. Once the recorder
indicates that the fraction is off the column, the collection tube should be detached
from the adapter.
5. The collection tube should then be connected to a tared 100 mL conical vial and
placed in a centrifuge tube (cotton padding around the collection tube and bottom of
the vial is recommended - refer to Figure 5-2). Centrifuge the assembly. The collected
GC fraction has now been collected. Detach the vial from the collection tube, cap the
vial and weight it.
GC Fraction
Collection Tube
12 mL
Centrifuge Tube
Cotton
Plug
100
μ
L
Conical Vial
Sample
Collects
Here
GC Collection Tube
GC Connection
Adapter
Vial,
Conical Reaction
(Threaded)
Figure 5-2
Microscale Prep Glassware
Содержание 400 Series
Страница 2: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 2 ...
Страница 4: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 4 ...
Страница 8: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 8 ...
Страница 34: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 34 ...
Страница 40: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 40 ...
Страница 56: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 56 ...
Страница 58: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 58 ...
Страница 60: ...GOW MAC Instrument Co 60 ...
Страница 62: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 62 ...
Страница 64: ...GOW MAC Instrument Co Series 400 402 Gas Chromatograph Operating Manual May 2017 Rev 7 64 ...