38
GMC-I Gossen-Metrawatt GmbH
(status: HOLD
→
HOLD)
Functions
!
The next memory location can be selected at any time from a
suspended
SEQUENCE
. The setpoints saved to the next
memory location are executed. Selected dwell times of the
exited and the new memory address are not taken into
consideration.
!
The device is switched from automatic sequence to step-by-
step control the first time this step is executed (see page 36).
•
The output retains its switching status (output On or Off).
•
When jumping from the stop address to the start address, the
remaining repetitions counter (rrep) remains unchanged.
Procedure:
☞
With
suspended
SEQUENCE,
– press the <FUNCTION> key and then browse with the
<
↑
> key, the <
↓
> key or the <FUNCTION> key
until SEq
step
appears
and execute the command with the
<ENTER>
key.
!
The
SEQUENCE
executes the next address.
!
Selected dwell times of the new and the exited memory
address are not taken into consideration.
!
The READY LED blinks slowly in the HOLD status.
Executing the Start Address and Beginning Step-by-Step Control
(status: HOLD
→
HOLD or RUN
→
HOLD)
Functions
!
The start address can be selected at any time from an
interrupted or a running
SEQUENCE
. The setpoints saved to the
start address are executed. Selected dwell times of the exited
and the new memory address are not taken into
consideration.
!
The device is switched from automatic sequence to step-by-
step control the first time this step is executed (see page 36).
•
The output is switched to the output On status.
•
The remaining repetitions counter is reset to the initial value
(rrep = rep).
Procedure:
☞
With
interrupted
or
running
SEQUENCE,
– press the <FUNCTION> key and then browse with the
<
↑
> key, the <
↓
> key or the <FUNCTION> key
until SEq
start
appears,
and execute the command with the
<ENTER>
key.
!
The
SEQUENCE
executes the start address.
!
Selected dwell times of the new and the exited memory
address are not taken into consideration.
!
The READY LED blinks slowly in the HOLD status.
4.7.4
buS – The “Interface” Function Group
The KONSTANTER can be set up for data transmission to a
computer in the interface configuration function group. In this
case, all of the settings previously selected with DIP switches are
entered via menu functions.
The serial port (RS 232) or the parallel port (IEEE 488) is used for
communication.
Addr – Selecting the Device Address (IEEE 488)
•
An address must be assigned to the SSP KONSTANTER in
order to assure that it functions correctly together with other
devices on the IEC bus.
•
If the KONSTANTER receives its own address from the computer, it
interprets subsequent data and executes the respective commands.
•
If it receives the address of another device, all subsequent
data are ignored until it receives its own address.
Settings
☞
Procedure
for selecting
Addr:
Settings.
☞
Press the <
FUNCTION
> key in order to access the
FUNCTION
menu.
!
The last edited function or function group is accessed
automatically.
☞
Select
bUS
configuration if necessary by pressing the
<
FUNCTION
> key, or the <
FUNCTION
> key and the <
↓
> key or
the <
↑
> key simultaneously.
!
bUS
appears at the left-hand display.
☞
Acknowledge by pressing the <ENTER> key.
!
The
bUS
function menu appears.
☞
Select the
Addr
function by pressing the <
FUNCTION
> key, or
the <
FUNCTION
> and the <
↓
> key or the <
↑
> key
simultaneously.
!
Addr
appears at the left-hand display, and the last selected
parameter appears at the right-hand display.
☞
Select an address for the SSP KONSTANTER with the <
↓
>
key or the <
↑
> key.
!
Any address within a range of 0 to 31 may be selected.
!
The selected address must be the same as the address
selected for the SSP KONSTANTER at the computer.
!
Address 0 is usually assigned to the controller.
!
Address 31 means unL (unlist).
☞
Acknowledge your selection with the <
ENTER
> key.
Figure 4.7.4 a
Path to Device Address Selection
From the bUS function group
Select the
device address
FUNCTION
Press repeatedly if necessary
ENTER
CE/LOCAL
FUNCTION
SAVE
RCL
Uset/Iset
FUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change
Содержание 62 N Series