GMC-I Gossen-Metrawatt GmbH
37
•
The output retains its switching status (output On or Off).
•
The
SEQUENCE
can be
resumed
(cont) at the next memory
location,
or ended at the current address
<
CE/LOCAL
> or
the stop
address
(stop) or
restarted
(go).
In addition,
step-by-step control
can be started at the
next
address
(Step) or at the
start address
(Start) as well.
Procedure:
☞
With
running
SEQUENCE,
– press the <
ENTER
> key or
– press the <FUNCTION> key and then browse with the
<
↑
> key, the <
↓
> key or the <FUNCTION> key
until SEq
hold
appears
and execute the command with the
<ENTER>
key.
!
The
SEQUENCE
stops
at the current address.
!
The READY LED blinks slowly in the HOLD status
.
Resuming the Sequence from the HOLD Status
(status: HOLD
→
RUN)
Functions
!
A suspended
SEQUENCE
can be resumed at any time.
•
The output retains its switching status (output On or Off).
The selected dwell time of the exited memory location is not
taken into consideration.
!
The sequence is now resumed starting with the next address,
and the values stored to the respective memory location (
Uset
,
Iset
and
tdef/tset
) are read out to the power output.
!
In the case of very long dwell times, you can advance
prematurely to the next memory location by
pressing the
<ENTER> key twice
(status: RUN
→
HOLD
→
RUN)
.
Procedure:
☞
With
suspended
SEQUENCE,
– press the <
ENTER
> key or
– press the <FUNCTION> key and then browse with the
<
↑
> key, the <
↓
> key or the <FUNCTION> key
until SEq
cont
appears
and execute the command with the
<ENTER>
key.
!
The
SEQUENCE
is resumed starting with the next memory
location.
!
The READY LED blinks rapidly in the Run status
.
Ending the Sequence at the Stop Address (status: HOLD
→
RDY)
Functions
!
A suspended
SEQUENCE
can be ended at any time by
executing the stop address.
•
If the memory location assigned to the stop address is empty,
the Uset and Iset values from the hold address are retained. If
the memory location assigned to the stop address is not
empty, Uset and Iset are changed to the corresponding
values. Selected dwell time Tset of the current memory
location is not taken into consideration.
•
The output status is determined by the contents of the stop
address. The output retains its switching status (output On or
Off) if the memory location assigned to the stop address is not
empty. The output is switched to the off status if the memory
location assigned to the stop address is empty.
Procedure:
☞
With
suspended
SEQUENCE,
– press the <FUNCTION> key and then browse with the
<
↑
> key, the <
↓
> key or the <FUNCTION> key
until SEq
stop
appears
and execute the command with the
<ENTER>
key.
!
The
SEQUENCE
is ended and the output status is determined
by the contents of the stop address.
!
The READY LED does not blink in the RDY status
.
Ending the Sequence at the Current Address (status: HOLD
→
RDY)
Functions
!
A suspended
SEQUENCE
can be ended at the current address
at any time.
•
Uset and Iset retain the values specified in the current memory
location. Selected dwell time Tset of the current memory
location is not taken into consideration.
•
The output retains its switching status (output On or Off).
Procedure:
☞
With
suspended
SEQUENCE,
– press the <CE/LOCAL> key.
!
The
SEQUENCE
is ended and the output status is determined
by the content of the current address.
!
The READY LED does not blink in the RDY status
.
Restarting the SEQUENCE
(status: HOLD
→
RUN or RUN
→
RUN)
Functions
!
A suspended or running
SEQUENCE
can be restarted at any
time.
•
The output is switched to the output On status.
!
In this case, all memory locations are executed, one after the
other, from the start address to the stop address, and the
respective values saved to each location (
Uset
,
Iset
and
tdef/
tset
) are read out to the power output and the signal output.
Procedure:
☞
With
interrupted
or
running
SEQUENCE,
– press the <FUNCTION> key and then browse with the
<
↑
> key, the <
↓
> key or the <FUNCTION> key
until SEq
go
appears
and execute the command with the
<ENTER>
key.
!
The
SEQUENCE
is restarted.
Figure 4.7.3 j
Suspending and Resuming a Sequence
Executing the Next Address and Beginning Step-by-Step Control
ENTER
ENTER
ENTER
SEQUENCE
SEQUENCE
Sequence is resumed
Resume sequence
Restart sequence
End sequence
Select:
The setting
is executed
in hold status
READY
READY
in run status
READY
Step-by-step, start address
Step-by-step, next address
FUNCTION
FUNCTION
CE/LOCAL
End at
current address