GMC-I Gossen-Metrawatt GmbH
67
Fixed response string length: 13 characters
Example (HP Basic):
OUTPUT 712;"IMIN?"
ENTER 712;A$
DISP A§
→
Display:
IMIN -000.010
Comment
→
IOUT? – Querying the Momentary Current Value
Functions
The IOUT function triggers measurement of the momentary
current value. The measured value is acquired over a period of
40 ms in a continuous fashion.
Syntax
IOUT?
Measuring Range
a) Manual Operation
b) Programming
Measured value query
Query command:
IO
UT
?
Response string:
IOUT
value
Response parameter format
value
:
±nnn.nnn
Fixed response string length: 13 characters
Example (HP Basic):
OUTPUT 712;"IOUT?"
ENTER 712;A$
DISP A§
→
Display:
IOUT +031.510
Comment
→
ISET, ISET? – Current Setpoint
Functions
ISET defines the output current setpoint value.
Syntax
ISET value
Parameter
Value
Parameter type: real number
The standard display format is: “XX.XX”. The current display
format is automatically changed to “X.XXX” for values< 10.00 A
for 12.5 A models, and to “XXX.X” for values as of 100 A for 150 A
models (manual setting resolution is reduced to 0.2 A as well).
The displayed setpoint is rounded off to the display format. Thus if
the smallest step size has been selected, minor setpoint
adjustments made with the keys or the rotary knob may not be
visible at the display.
a) Manual Operation
b) Programming
Setting
Setting command:
IS
ET
value
Example (HP Basic):
OUTPUT 712;"ISET 11.3"
!Current setpoint: 11.3 A
Setting query
Query command:
IS
ET
?
Response string:
ISET
value
Parameter format
value
:
+nnn.nnn
Fixed response string length: 13 characters
Example (HP Basic):
OUTPUT 712;"ISET?"
ENTER 712;A$
DISP A$
→
Display:
ISET +011.300
Comments
If an attempt is made to set ISET to a value higher than ILIM,
the setting command is not executed and bit 1 is set in event
register B (limit error), as is bit 4 in the standard event register
(execution error).
MINMAX, MINMAX? – Min-Max Storage for Measured U and I
Values
Functions
The MINMAX function saves minimum and maximum measured
voltage and current values to memory. The saved values, UMIN,
UMAX, IMIN and IMAX, can then be displayed or queried via the
interfaces.
Syntax
MINMAX
status
Parameter
Status
One of the following text parameters can be selected:
OFF
Storage of Min-Max values deactivated
ON
Storage of Min-Max values activated
RST
Contents of Min-Max memory are reset, i.e. are replaced
with the momentary measured value for the
corresponding parameter:
Umin = Uout
Umax = Uout
Imin = Iout
Imax = Iout
Default setting after RESET (*RST):
OFF
Device
Type
Measuring Range
Resolution
Nom.
Current [A]
Min.
a)
[A]
a. Measured values which violate the measuring range lower limit are identified
at the display as “–OL”, and in the data string as “–999999.”.
Max.
b)
[A]
b. Measured values which exceed the measuring range upper limit are
identified at the display as “+OL”, and in the data string as “+999999.”.
Remote
[mA]
Display
[mA]
12.5
–0.478
13.342
2
2 / 10
25
–0.960
26.685
5
5 / 10
50
–1.92
53.370
10
10
75
–2.880
80.060
10
10
100
–3.840
106.740
20
20 / 100
150
–5.760
160.120
20
20 / 100
Device
Type
Setting Range
Step Size
Nom.
Current [A]
min.
[A]
max.
a)
[A]
Remote
b)
[A]
Manual
c)
[A]
12.5
0.00
12.50
0.003125
0.003125/0.0125/0.1
25
0.00
25.00
0.00625
0.00625/0.1/1.0
50
0.00
50.00
0.0125
0.0125/0.1/1.0
75
0.00
75.00
0.02
0.02/0.1/1.0
100
0.00
100.0
0.025
0.025/0.1/1.0
150
0.00
150.0
0.04
0.04/0.2/1.0
Default setting after RESET (*RST): 0.00 [A]
a. However, ISET cannot be set to a value of greater than the momentary current limit value ILIM.
b. If the selected value is not a whole number multiple of the step size, it is rounded off accordingly.
If necessary, the numeric value is rounded off once again for the 4-place digital display.
c. Can be selected by (repeatedly) pressing the <RESOLUTION> key. The blinking decimal place at
the display indicates the selected step size.