PART ONE - INTRODUCTION
SECTION 2 - SPECIFICATIONS
Rev. 0-201
PRO-series Model C3 Conductivity Transmitter
15
S
ECTION
2
2.1
Operational
Display....................................... Two-line by 16 character LCD
NOTE:
The measured value (conductivity, resisitivity or TDS) and temperature
can be displayed separately or shown together on a single screen.
The corresponding 4-20 mA analog output value can also be shown.
Measurement
Ranges
Conductivity .......................... µS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000
mS/cm: 0-2.000, 0-20.00 or 0-200.0
Resistivity ............................. 0-19.99 M
Ω
•
cm or 0-999.9 K
Ω
•
cm
TDS ...................................... 0-9999 ppm or 0-9999 ppb
Temperature ......................... -4.0 to +392.0
°
F or -20.0 to +200.0
°
C
Analog Output....................... 4.00-20.00 mA
Ambient Conditions:
Operation.............................. -4 to +140
°
F (-20 to +60
°
C); 0-95% relative
humidity, non-condensing
Storage ................................. -22 to +158
°
F (-30 to +70
°
C); 0-95% relative
humidity, non-condensing
Temperature Compensation ....... Automatic from -4.0 to +392.0
°
F (-20.0 to
+200.0
°
C) with selection for Pt 1000 ohm
RTD or Pt 100 ohm RTD temperature
element, or manually fixed at a user-entered
temperature
NOTE:
The selected measurement (conductivity, resistivity or TDS)
determines which of the following temperature compensation
methods are available:
Linear % per
°
C slope, built-in ammonia temp. properties table,
built-in natural water temp. properties table, or no compensation
Sensor-to-Transmitter Distance .. 300 ft. (91 m) maximum
Power Requirements:
Two-wire Hookup .................. 16-30 VDC
Three-wire Hookup ................ 14-30 VDC (16 VDC min. with RS-485 comm.)
Four-wire Hookup.................. 12-30 VDC (16 VDC min. with RS-485 comm.)
Calibration Methods:
Sensor ZERO........................ With the dry sensor in air, press keys to
(all measurements)
initiate automatic system zeroing
Conductivity Measurement:
DRY-CAL......................... Enter GLI-certified cell constant “K” value and
temperature “T” factor of sensor (during
initial transmitter configuration)
1-POINT SAMPLE ........... Enter one reference solution value or one
sample value (determined by laboratory
analysis or a comparison reading)
Resistivity Measurement:
DRY-CAL......................... See description above (for conductivity)
TDS Measurement:
DRY-CAL......................... See description above (for conductivity)
1-POINT SAMPLE ........... See description above (for conductivity)
SPECIFICATIONS