PART THREE - OPERATION
SECTION 5 - TEST/MAINTENANCE
Rev. 0-201
PRO-series Model C3 Conductivity Transmitter
61
5.6 SIM SENSOR Setting
☞
5.7 RESET CONFIGURE
Values to Factory
Defaults
☞
•
SENSOR:
Depending on the configured
measurement, selects the simulated value to be
a conductivity, resistivity or TDS value.
•
TEMPERATURE:
Selects the simulated value to
be a temperature value.
3.
With the desired choice displayed, press
ENTER key
to
enter this selection and return to the previous level of
the TEST/MAINT menu branch.
After selecting the type of simulated measurement (subsec-
tion 5.5), set the desired simulation value.
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display a screen like
.
NOTE:
The analog output signal is now active. It has a
mA value that corresponds to the measurement
value shown on this screen.
3.
Use
arrow keys
to adjust the displayed simulation
value to the desired value.
4.
To remove the simulated output and return to previous
level of TEST/MAINT menu branch, press
ESC key
or
ENTER key
.
You can conveniently reset all stored configuration settings
at the same time --
but not calibration settings
-- to their
factory-set defaults shown in Table B.
NOTE:
Resetting configuration values also excludes the
SELECT MEASURE function (conductivity, resistiv-
ity or TDS) which remains as is until you change it.