PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model C3 Conductivity Transmitter
Rev. 0-201
38
CONFIG LINEAR
Temperature Compensation
(configuration not needed for
other compensation methods)
4.
With the desired choice displayed, press
ENTER key
to
enter this selection. If the “NaCl” conversion factor was
selected, TDS measurement configuration is complete.
If you selected “USER DEFINED,” you must enter a
conductivity-to-TDS conversion factor:
A.
With the
screen displayed,
press
Ø
Ø
key
once to display
.
B.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the
displayed value to the desired conductivity-to-TDS
conversion factor, and press
ENTER key
to enter
the value.
C. After the
screen re-appears,
press
ESC key
once to return to the
screen.
Only when LINEAR is the selected temperature com-
pensation, determine if the transmitter should be further
configured by entering a specific slope (% per
°
C), and
a reference temperature.
If the built-in AMMONIA or
NATURAL WATER properties table or NONE was selected,
disregard this subsection -- no compensation configuration
is needed.
LINEAR compensation factory defaults are 2.00%/
°
C slope
and 25.0
°
reference temperature.
These values are ap-
propriate for most aqueous solutions.
Use chemical
handbook tables to find values for uncommon solutions. To
enter different values, do the following:
1.
With the
or
screen displayed, press
Ø
Ø
key
until
appears.