PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model C3 Conductivity Transmitter
Rev. 0-201
48
Table B -- TRANSMITTER CONFIGURATION SETTINGS (RANGES/CHOICES and DEFAULTS -- continued)
Displayed Screen Title
Entry Range or Choices (where applicable)
Factory Default
Your Setting
TEMPERATURE Display Configuration Setting
CONFIGURE:
°
C OR
°
F?
°
C or
°
F
°
C
OUTPUT Configuration Settings
OUTPUT:
SET PARAMETER?
SENSOR or TEMPERATURE
SENSOR
OUTPUT:
SET 4mA VALUE?
CONDUCTIVITY:
µS/cm:
0-2.000, 0-20.00, 0-200.0 or 0-2000
mS/cm: 0-2.000, 0-20.00 or 0-200.0
RESIST: 0-19.99 M
Ω
•
cm or 0-999.9 K
Ω
•
cm
TDS: 0-9999 ppm or 0-9999 ppb
TEMP: -20.0 to +200.0
°
C or -4.0 to 392.0
°
F
CONDUCTIVITY:
0 µS/cm or 0 mS/cm
RESIST: 0 M
Ω
or 0 K
Ω
•
cm
TDS: 0 ppm or 0 ppb
TEMP: 0.0
°
C or 32.0
°
F
OUTPUT:
SET 20mA VALUE?
CONDUCTIVITY:
µS/cm:
0-2.000, 0-20.00, 0-200.0 or 0-2000
mS/cm: 0-2.000, 0-20.00 or 0-200.0
RESIST: 0-19.99 M
Ω
•
cm or 0-999.9 K
Ω
•
cm
TDS: 0-9999 ppm or 0-9999 ppb
TEMP: -20.0 to +200.0
°
C or -4.0 to 392.0
°
F
CONDUCTIVITY:
µS/cm: 2.000, 20.00, 200.0,
or 2000
mS/cm: 2.000, 20.00 or
200.0
RESIST: 19.99 M
Ω
or 999.9 K
Ω
•
cm
TDS: 9999 ppm or 9999 ppb
TEMP: 200.0
°
C or 392.0
°
F
OUTPUT:
SET FILTER?
0-60 seconds
0 seconds
OUTPUT:
SET FAIL LEVEL?
OFF, 4mA or 20 mA
OFF
PASSCODE Configuration Setting
SET PASSCODE?
DISABLED or ENABLED
DISABLED
TEST/MAINT Simulation Function Settings
SELECT SIM?
SENSOR or TEMPERATURE
SENSOR
SIM SENSOR?
CONDUCTIVITY:
µS/cm:
0-2.000, 0-20.00, 0-200.0 or 0-2000
mS/cm: 0-2.000, 0-20.00 or 0-200.0
RESIST: 0-19.99 M
Ω
•
cm or 0-999.9 K
Ω
•
cm
TDS: 0-9999 ppm or 0-9999 ppb
TEMP: -20.0 to +200.0
°
C or -4.0 to 392.0
°
F
Present measured value of
sensor’s selected parameter