PART TWO - INSTALLATION
SECTION 3 - ELECTRICAL CONNECTIONS
Rev. 0-201
PRO-series Model C3 Conductivity Transmitter
21
S
ECTION
3
☞
☞
Figure 2-4 shows the terminal block arrangement and
terminal designations for the transmitter.
NOTE:
All terminals are suitable for single wires up to 14
AWG (2.5 mm
2
).
Wiring Tip!
To comply with European Community (CE) electromagnetic
compatibility requirements, follow these general wiring guidelines:
1. Locate transmitter as far as possible from motors and other non-
CE certified devices with excessive electromagnetic emissions.
2. Use GLI-specified ferrites and cables. Failure to do so may elimi-
nate compliance. Locate all ferrites as close as possible to the
transmitter.
3. Loop these cables through the specified ferrite using the required
number of loops:
◆
DC Power Supply Cable (GLI 1W0980 two-conductor plus
shield): Connect cable shield to earth ground at the supply
end. Loop cable 4 times through ferrite (Fair-Rite Corp.
#2643102002 or equivalent).
◆
Sensor Cable: Keep cable shield as short as possible and
connect it to the SHIELD terminal. Loop cable 3-1/2 times
through ferrite (Fair-Rite Corp. #2643804502 or equivalent).
◆
Analog mA Output Cable (four-wire hookup only -- GLI
1W0980 two-conductor plus shield): Connect cable shield to
earth ground at the supply end. Loop cable 4 times through
ferrite (Fair-Rite Corp. #2643102002 or equivalent).
TB1
TB2
PRO-C-ESP: 2E Conductivity
Shield
(Ground)
White
(Temp)
Blue
(Signal)
Red
(Drive)
(S
ee
ma
nua
l for
3,
4 W
ire
C
onn
ect
ion
s)
S
e
rial
#:
47
47
47
47
47
9
2-
Wi
re C
o
n
nec
tion
3
(-)
6
7
5
4
14
-30
VD
C
(+)
1
2
3
6
NC
Black
7
5
4
1
NC
2
(Ground)
FIGURE 2-4 Transmitter Terminal Designations
ELECTRICAL CONNECTIONS