PART THREE - OPERATION
SECTION 4 - TRANSMITTER CALIBRATION
Rev. 0-201
PRO-series Model C3 Conductivity Transmitter
55
☞
Remove any bubbles clinging to the
outer and center electrodes by gently
stirring the sensor in the solution. Sen-
sors with a deeper recessed center
electrode require more vigorous stirring.
2.
Press
MENU key
to display a “MAIN MENU” screen.
If the
screen is not showing, use
Ø
Ø
or
×
×
key
to display it.
3.
Press
ENTER key
to display
.
4.
Press
ENTER key
again to display
.
5.
Press
ENTER key
again to display
.
With the sensor in the process (or conductivity refer-
ence solution), press
ENTER key
to confirm. This
active
screen appears showing the
measurement reading.
NOTE:
During calibration, the analog output is auto-
matically “held” at the last measured value.
6.
Wait for the reading to stabilize which may take a few
minutes. Then press
ENTER key
. The “PLEASE WAIT”
screen may appear if the reading is still too unstable.
After the reading has stabilized, this static
screen appears showing the “last
measured” value.
7.
Use
arrow keys
to adjust the displayed value to
exactly match the known value of the process sample
(or reference solution).
8.
Press
ENTER key
to enter the value and complete cali-
bration (“CONFIRM CAL OK?” screen appears).
9.
If the sensor was immersed in a reference solution, re-
install the sensor into the process.
10. Press
ENTER key
to display the active measurement
reading on the “1 POINT SAMPLE: CONFIRM AC-