PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
Rev. 0-201
PRO-series Model C3 Conductivity Transmitter
41
Select
TEMP ELEMENT
Type
☞
6.
After the
screen re-appears, press
ESC key
once to return to the
screen.
Configure the transmitter for either automatic temperature
compensation (uses Pt 1000 ohm RTD built into sensor), or
fixed MANUAL temperature compensation. When using
MANUAL you must determine and enter a specific tem-
perature value.
NOTE:
When a temperature element type has been se-
lected but the element is not connected to the
transmitter, a “WARNING: CHECK STATUS” mes-
sage will appear. To prevent or clear the message,
connect the element or select “MANUAL.”
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
.
3.
Press
ENTER key
again to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices:
•
PT1000:
Configures transmitter for use with a
Pt 1000 RTD temperature element (used in all GLI
Model 3400-series contacting conductivity sensors).
•
MANUAL:
Configures transmitter for fixed manual
temperature compensation when not using a
temperature element.
4.
With the desired choice displayed, press
ENTER key
to
enter this selection. When “MANUAL” is selected, you
must set the specific manual temperature compensation
value: