PRO-series Model C3 Conductivity Transmitter
Rev. 0-201
58
S
ECTION
5
☞
5.1 STATUS Checking
(transmitter and
sensor)
The transmitter has TEST/MAINT menu screens to:
•
Check system status for the transmitter and sensor.
•
Hold analog output at its last measured value.
•
Provide analog output test signal to confirm operation of
connected device.
•
Identify transmitter firmware EPROM version.
•
Simulate a measurement or temperature signal to
exercise the measurement loop.
•
Reset configuration (not calibration) values to defaults.
•
Reset calibration (not configuration) values to defaults.
NOTE:
When the passcode feature is enabled (Section
3.5), you must successfully enter the passcode
before attempting to use the TEST/MAINT menu
screens.
The system diagnostic capabilities of the transmitter enable
you to check the operating status of the transmitter and
sensor. The MEASURE screen will flash the “WARNING
CHECK STATUS” message when a sensor or transmitter
“FAIL” diagnostic condition has been detected. To deter-
mine the condition causing the warning, display the “STATUS”
screens.
1.
Press
MENU key
to display a “MAIN MENU” screen.
If the
screen is not showing, use
Ø
Ø
or
×
×
key
to display it.
2.
Press
ENTER key
to display
.
3.
Press
ENTER key
again to display “STATUS: ANALYZER
OK” screen. This screen confirms that the transmitter is
operating properly. If “FAIL” appears, it may mean:
•
Analog-to-digital converter not responding.
•
Internal serial communications failure.
TEST/MAINTENANCE