E44.X.01.6C-04
Mounting- and Operating Manual GIR 300
page 7 of 34
4
Display and operating elements
4.1 Display elements
Main display:
Display element to show minimum / maximum and measuring value.
It also displays errors and parameters.
LED 1
:
Indicates the state of output 1
LED 2
:
Indicates the state of output 2
4.2 Operating elements
Button 1:
activates menu "Switching points and alarm boundaries"
Button 1 + 5, >2s: activates menu "Configuration of Output functions"
Menu:
save value or step
to next parameter
Button 2:
display max. value
button 2 + 5, >2s: activates menu „Select input signal“
button 2 + 3, >2s: reset min-/max. value
Menu: press short
= increase value
press long
= roll-function with overflow-function *)
Button 3:
display min. value
button 3 + 5, >2s: activates menu
„Offset- and slope adjustment“
button 3 + 2, >2s: reset min-/max. value
Menu: press short
= decrease value
press long
= roll-function with overflow-function *)
*) The input is made with the buttons 2 and 3. When pressing the button once the value will be raised (but-
ton 2) by one or lowered (button 3) by one.
When holding the button pressed for longer than 1 sec. the value starts counting up or down, the count-
ing speed will be raised after a short period of time.
The device also features a ‘overflow-function‘, when
reaching the upper limit of the range, the device switches to the lower limit, vice versa.