E44.X.01.6C-04
Mounting- and Operating Manual GIR 300
page 19 of 34
- When
pressing button 1 again, the display will show “
Fr.Hi
“ (frequency high
= upper frequency
range limit
).
- Use button 2 and button 3 to select the highest frequency that may occur when measuring.
- Press button 1 to validate your selection. The display shows
“
Fr.Hi
“ again.
-
When pressing button 1 again, the display will show “
dP
“ (decimal point).
- Use button 2 and button 3 to select the desired decimal point position.
-
Press button 1 to validate your selection. The display shows “
dP
“ again.
- When pressing
button 1 again, the display will show “
di.Lo
“ (display low
= display at lower frequency
range limit
).
- Set the value the device shall display at the lower frequency range limit by pressing button 2 or
button 3.
- Press button 1 to validate your selection
. The display shows “
di.Lo
“ again.
-
When pressing button 1 again, the display will show “
di.Hi
“ (display high
= display at upper frequency
range limit
).
- Set the value the device shall display at the upper frequency range limit by pressing button 2 or
button 3.
-
Press button 1 to validate your selection. The display shows “
di.Hi
“ again.
-
When pressing button 1 again, the display will show “
Li
“ (limit
= measuring range limitation
).
- Use button 2 and button 3 to select the desired measuring range limitation.
Display
Measuring range limit
Notes
off
Inactive
Exceeding of the measuring-frequency is
tolerable until you reach the maximum
measuring range limit.
on.Er
Active,
(displays error)
The measuring range is exactly bounded
by the selected frequency-measuring-
range-limit. When exceeding or shortfal-
ling of the limit the device will display an
error message.
on.rG
Active,
(displays the selected limit)
The measuring range is exactly bounded
by the selected frequency-measuring-
range-limit. When exceeding or shortfal-
ling of the limit the device will display the
lower or upper display-range-limit.
[e.g. for humidity: when shortfalling or
exceeding the device will display 0% or
100%]]
Hint:
When exceeding the maximum range limit (10 kHz) independently from the limit setting an error message
will be displayed (“Err.1“).
-
Press button 1 to validate your selection. The display shows “
Li
“ again.
- When
pressing button 1 again, the display will show “
FiLt
“ (Filter
= digital filter
).
- Use button 2 and button 3 to select the desired filter value [in sec.].
Usable values: 0.01 ... 2.00 sec.
Explanation: this digital filter is a digital replica of a low pass filter.
-
Press button 1 to validate your selection. The display shows “
FiLt
“ again.
Now your device is adjusted to your signal source. The only thing you left do is to adjust the outputs of the
device.
- When pressing button 1 again, the disp
lay will show “
outP
“. (Output)
For configuring the outputs of the device, please follow the instructions shown in chapter 6.9.
DP
FR.K,
D,.LO
D,.K,
L,
F,LT
OVTP