Chapter 5: System Maintenance
The ORS SITEPRO™ 2000 is designed for trouble free operation with minimal maintenance required. The following
simple maintenance tasks should be carried out at the specified intervals.
Vacuum Enclosure
At yearly intervals, the insides of the enclosure should be vacuumed to remove accumulated dust.
Clean Window
As needed, clean the enclosure window using a solution of soap and warm water.
Change Desiccant
If desiccant is being used to control moisture accumulation inside the enclosure, this material should be renewed on a
monthly basis or at the intervals recommended by the desiccant manufacturer.
Clean Probes
It is essential that the probe shafts and floats be cleaned on a regular basis. Use Alconox detergent, warm water and a
soft brush. The required frequency of cleanings is highly site specific and must be determined by the user.
Glass cleaners containing ammonia will damage the enclosures Lexan windows. Also,
prevent hydrocarbons from contacting the window.
Geotech does not supply desiccant.
Failure to clean the probe will result in fouling that could cause system
malfunction.