FIR (Fc @
-
3 dB) Filter (Digital Anti
-
Alias)
Passband
Stopband
δp : Passband ripple
δs : Stopband attenuation
ωp
ωc
ωs
: Passband frequency
: Corner frequency
: Stopband frequency
-3 dB
Frequency [kHz]
Magnitude [dB]
+ δp
- δp
δs
ωp
ωs
ωc
Figure L.7:
Digital FIR (Fc @
-
3 dB) filter
When FIR (Fc @
-
3 dB) filter is selected, this is always a combination of an analog Butterworth anti
-
alias filter and a digital FIR (Fc @
-
3 dB)
filter. The FIR (Fc @
-
3 dB) is an adapted FIR filter. Its ωp is reduced by a factor of ≈1.4 compared to the FIR (Fc @
-
0.1 dB) filter. It is supported
by Perception V6.40 and higher.
Analog anti
-
aliasing filter bandwidth
370 kHz
±
20 kHz (
-
3 dB)
Analog anti
-
aliasing filter characteristic
7
-
pole Butterworth, extended passband response
FIR (Fc @
-
3 dB) filter characteristic
12
-
pole FIR; FIR is a purely digital characteristic. Its closest analog resemblance is to an
Elliptic filter. However, FIR has both overshoot and pre
-
shoot on the step response. This
means that the ringing on the signal starts before the step input is started and that the ringing
continues after the step input is complete.
FIR (Fc @
-
3 dB) filter user selection
Auto tracking for sample rate divided by: 4, 10, 20, 40
FIR (Fc @
-
3 dB) filter bandwidth (ωc)
Auto tracking the sample rate with the selected FIR (Fc @
-
3 dB) filter; user selection from
0.031 Hz to 50 kHz
FIR (Fc @
-
3 dB) filter passband flatness (ωp)
(1)
0.1 dB; DC to ≈ωc/1.4 (Adapted FIR filter behavior)
FIR (Fc @
-
3 dB) filter stopband attenuation (δs)
60 dB
FIR (Fc @
-
3 dB) filter roll
-
off
72 dB/Octave
FIR (Fc @ -3 dB) 50 kHz Overview
10
0
-10
-20
-30
-40
-50
-60
-70
-80
Frequency [kHz]
Ma
gni
tu
de
[dB
]
Passband flatness
Frequency [kHz]
Ma
gni
tu
de
[dB
]
0.2
0.1
0.05
0.15
-0.1
-0.15
-0.05
-0.2
0
10000
1000
100
10
1
0.1
1000
100
10
1
0.1
Figure L.8:
Representative FIR examples
(1)
Measured using a Fluke 5700 calibrator, DC normalized
GEN3t
990
I3995-3.1 en HBM: public
Содержание GEN3t
Страница 1: ...Portable Data Recorder GEN3t English User Manual I3995 3 1 en HBM public...
Страница 4: ...4 I3995 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3t I3995 3 1 en HBM public 51...
Страница 1122: ...1122 I3995 3 1 en HBM public...
Страница 1123: ......