
Bridge completion
-
full (4/4) bridge
A full bridge type sensor is a sensor that has all four bridge resistors on
-
board;
no completion is required.
A
B
C
D
+V_EXC
−V_EXC
+AMP INPUT
−AMP INPUT
R1
R2
R3
R4
BRIDGE RESISTOR
BRIDGE COMPLETION
Figure 12.40:
Full
-
bridge layout
At least four wires are necessary to connect a bridge such as this. Please refer
to Figure 12.37 for connection details. Inform the Perception software when a
full bridge is used.
To select full bridge completion in Perception
To select full bridge completion, proceed as follows:
1
In Perception, go to the
Settings
sheet.
2
In the task pane, select the
Bridge
in the
Input
section.
3
Select one or more channels.
GEN3t
298
I3995-3.1 en HBM: public
Содержание GEN3t
Страница 1: ...Portable Data Recorder GEN3t English User Manual I3995 3 1 en HBM public...
Страница 4: ...4 I3995 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3t I3995 3 1 en HBM public 51...
Страница 1122: ...1122 I3995 3 1 en HBM public...
Страница 1123: ......