The first interval timer can be compared to the trigger hold off feature described
above. The second interval timer defines a period in which a trigger event must
occur. If not, it is not a related trigger event.
Interval timer
-
NotBetween
The inverse function of the Between mode of the interval timer is the
NotBetween mode. The second interval is not used to define a trigger
-
safe area,
but to denote a trigger
-
restricted area. A trigger event within the first interval is
valid. A trigger event within the second interval resets the trigger logic. A trigger
is also generated when both interval timers expire. A typical use of this mode
is to detect changes in the interval between pulses ("too early" / "too late"
detection).
P
1
6
8
7
2 3
5
9
4
10
A
A
Figure 10.16:
Interval timer
-
NotBetween
The following sequence explains how this mode functions:
1
The first trigger event starts Interval Timer 1.
2
Interval 1
3
If a trigger event occurs within the first interval, a trigger is generated.
4
End of sweep.
5
The first new trigger event starts Interval Timer 1.
6
Interval Timer 1 expires and Interval Timer 2 is started.
7
A trigger event occurs within the second interval. Interval Timer 1 is
restarted.
8
Interval 2
9
Interval Timer 1 expires and Interval Timer 2 is started.
10
Interval Timer 2 expires and a trigger is generated.
A
Trigger
GEN3t
150
I3995-3.1 en HBM: public
Содержание GEN3t
Страница 1: ...Portable Data Recorder GEN3t English User Manual I3995 3 1 en HBM public...
Страница 4: ...4 I3995 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3t I3995 3 1 en HBM public 51...
Страница 1122: ...1122 I3995 3 1 en HBM public...
Страница 1123: ......