Wideband Filter (No Anti
-
Alias Protection)
When wideband is selected, there is neither an analog anti
-
alias filter nor any digital filter in the signal path. Therefore, there is no anti
-
alias
protection when wideband is selected.
Bandwidth
Between 120 kHz and 160 kHz (
-
3 dB) all ranges < 100mV
Between 450 kHz and 520 kHz (
-
3 dB) all ranges ≥ 100 mV
0.1 dB passband flatness
(1)
DC to 10 kHz all ranges < 100 mV
DC to 80 kHz all ranges ≥ 100 mV
Magnitude [
dB]
-150
-130
-110
-90
-70
-50
-30
-10
10
Magn
itu
de [
d
B]
-150
-130
-110
-90
-70
-50
-30
-10
10
Frequency [kHz]
10000
1000
100
10
1
0.1
Frequency [kHz]
10000
1000
100
10
1
0.1
1000
100
10
1
0.1
Frequency [kHz]
1000
100
10
1
0.1
Frequency [kHz]
± 20 mV
Wideband passband flatness
±1 V
Wideband passband flatness
± 20 mV
Wideband overview
±1 V
Wideband overview
Magnitude [
dB]
Magn
itu
de [
d
B]
Magnitude [
%]
2.33
1.74
1.15
0.57
-0.57
-1.14
-1.71
-2.28
-0.2
-0.15
-0.1
-0.05
0.05
0.1
0.15
0.2
0
0
0
Magn
itu
de [
%]
2.33
1.74
1.15
0.57
-0.57
-1.14
-1.71
-2.28
-0.2
-0.15
-0.1
-0.05
0.05
0.1
0.15
0.2
0
Figure A.43:
Representative Wideband examples
(1)
Measured using a Fluke 5700 calibrator, DC normalized
GEN3t
I3995-3.1 en HBM: public
465
Содержание GEN3t
Страница 1: ...Portable Data Recorder GEN3t English User Manual I3995 3 1 en HBM public...
Страница 4: ...4 I3995 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3t I3995 3 1 en HBM public 51...
Страница 1122: ...1122 I3995 3 1 en HBM public...
Страница 1123: ......