CAT IV
: This category is for measurements directly connected to the source of
a low voltage mains installation. Measurements for this category are
overcurrent protection devices, ripple control units, etc. This category expects
that there is a minimum of one level of overcurrent protection between the
transformer and connection point of the measurement circuit. (See Figure 2.1).
Figure 2.1:
Category indication in accordance with IEC 61010
-
2
-
030:2010
Example:
A measurement device is specified as 600 V CAT II, maximum input
voltage 1000 V DC.
Table 2.1: Insulation test voltages in accordance with IEC 61010
-
2
-
030:2010
Nominal Voltage
(V RMS or V DC)
IEC 61010
-
2
-
030:2010
5 sec. AC test (V RMS)
Impulse test (V)
CAT II
CAT III
CAT IV CAT II
CAT III
CAT IV
≤ 150
840
1390
2210
1550
2500
4000
> 150 ≤ 300
1390
2210
3310
2500
4000
6000
> 300 ≤ 600
2210
3310
4260
4000
6000
8000
> 600 ≤ 1 000
3310
4260
6600
6000
8000
12000
Using the table above, it can be concluded that this specification informs the
user that the device passed the insulation tests; 5 sec at 2210 V RMS and
impulse 4000 V. The maximum operating input voltage is 1000 V DC. This
device is to be used to measure CAT II circuitry up to 600 V.
GEN3t
I3995-3.1 en HBM: public
29