Analog Input Section
Input overload protection
Overvoltage impedance change
The activation of the overvoltage protection system results in a reduced input impedance.
The overvoltage protection is not active for as long as the input voltage remains less than
200% of the selected input range or 1250 V, whichever value is the smallest.
Maximum nondestructive voltage
±
2000 V DC
Maximum overload without auto range
200% of selected range
Automatic auto range
When overload causes the amplifier to overheat, the amplifier increases its range in steps
of a factor of 10 until the overload ceases. When the overload exceeds 1000 V, the input
signal is disconnected and the amplifier input is grounded. When the temperature returns
to normal, the range that was originally selected is restored. The automatic auto range
cannot be turned off.
Overload recovery time
Restored to 0.1% accuracy in less than 5 μs after 200% overload
Maximum Static Error (MSE)
Maximum static error (MSE)
Typical
Maximum
All filters
0.02% of Full Scale
0.05% of Full Scale
±
10 μV
Ranges [V]
Ranges [V]
Detail MSE (all filters)
MSE (all filters)
Typical
Maximum
All filters
Typical
[%]
Maximum
[%]
± 0.01 V
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.020
0.100
0.075
0.060
0.055
0.053
0.051
0.051
0.051
0.051
0.051
0.051
0.051
0.051
0.051
0.051
0.051
± 0.02 V
± 0.05 V
± 0.1 V
± 0.2 V
± 0.5 V
± 1 V
± 2 V
± 5 V
± 10 V
± 20 V
± 50 V
± 100 V
± 200 V
± 500 V
± 1000 V
Er
ro
r
[%
]
Er
ro
r
[%
]
0
0.02
0.04
0.06
0.08
0.1
0.12
Typical
Maximum
0
0.01
0.02
0.03
0.04
0.05
0.06
±
0.
0
1
±
0.
0
2
±
0.
0
5
±
0.
1
±
0.
2
±
0.
5
±
1
±
2
±
5
±
10
±
20
±
50
±
100
±
200
±
500
±
1000
Figure A.154:
Maximum static error (MSE)
GEN3t
I3995-3.1 en HBM: public
613