12.9
GN401 Optical Fiber Isolated 100 MS/s input card
l
4 transmitters per receiver
card
l
Digital fiber optic connection,
noise/error and drift free
l
Cable length up to 800 m
l
Automatic cable length phase
compensation
l
1 isolated, unbalanced
differential input per
transmitter
l
Battery powered transmitter
l
1.8 kV RMS
isolation
continuous powered transmitter
l
±
20 mV to
±
100 V input ranges
l
Analog/digital anti
-
alias filters
l
15 or 14 bit resolution
l
Metal BNC input on transmitter
The optical fiber isolated system consists of up to four transmitter units (GN110,
GN111, GN112 or GN113) connected to the GN401 receiver card built into any
GEN series mainframe using a digital fiber optic cable.
By converting the analog signal into a digital signal and transmitting the signal
to the receiver card via fiber optic cable, the transmission does not add any drift
or error to the measured signal. The automatic cable length compensation
phase matches all fiber optic isolated channels to any standard analog input
channel. The GN112 and GN113 offer continuous powered isolation at 1.8 kV
RMS, while the GN110 and GN111 offer higher isolation options using battery
power with an continuous operation time of 30 hours. Optimum anti
-
alias
protection is achieved by the 6
-
pole analog anti
-
alias filter combined with a fixed
sample rate Analog
-
to
-
Digital converter. At lower sample rates the digital anti
-
alias filters allow for a large range of 8
th
order Bessel IIR filters with precise
phase match and ultra low noise output.
The GN112 and GN113 offer continuous powered isolation at 1.8 kV RMS,
while the GN110 and GN111 offer higher isolation options using battery power
with an continuous operation time of 30 hours.
For more information on the GN401 Optical Fiber Isolated 100 MS/s input card,
please refer to "B2629–2.1 en (GEN series GN401)" on page 423.
GEN3t
I3995-3.1 en HBM: public
333
Содержание GEN3t
Страница 1: ...Portable Data Recorder GEN3t English User Manual I3995 3 1 en HBM public...
Страница 4: ...4 I3995 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3t I3995 3 1 en HBM public 51...
Страница 1122: ...1122 I3995 3 1 en HBM public...
Страница 1123: ......