Wideband (No Anti
-
Alias Protection)
When wideband is selected, there is neither an analog anti
-
alias filter nor any digital filter in the signal path. Therefore, there is no anti
-
alias
protection when wideband is selected.
Wideband bandwidth
Between 26 MHz and 33 MHz (
-
3 dB)
0.1 dB passband flatness
(1)
DC to 1 MHz
- 0.2
- 0.15
-0.1
-0.05
0
0.05
0.1
0.15
0.2
0.001
0.001
0.01
0.1
1
10
100
0.01
0.1
1
10
100
1000
-150
-130
-110
-90
-70
-50
-30
-10
10
Frequency [MHz]
Frequency [MHz]
±
2 V: Wideband passband flatness
±
2 V: Wideband overview
M
a
g
n
it
u
d
e
[dB]
M
a
g
n
it
u
d
e
[dB]
0
M
a
g
n
it
u
d
e
[
%
]
2.33
1.74
1.15
0.57
-0.57
-1.14
-1.71
-2.28
±
20 V: Wideband passband flatness
0.001
0.01
0.1
1
10
100
1000
-150
-130
-110
-90
-70
-50
-30
-10
10
Frequency [MHz]
Frequency [MHz]
±
20 V: Wideband overview
0.001
0.01
0.1
1
10
100
-0.2
-0.15
-0.1
-0.05
0
0.05
0.1
0.15
0.2
M
a
g
n
it
u
d
e
[dB]
M
a
g
n
it
u
d
e
[dB]
M
a
g
n
it
u
d
e
[
%
]
2.33
1.74
1.15
0.57
0
-0.57
-1.14
-1.71
-2.28
Figure A.53:
Representative Wideband examples
(1)
Measured using Fluke 5700A and 5820A calibrators, DC normalized
GEN3t
I3995-3.1 en HBM: public
481
Содержание GEN3t
Страница 1: ...Portable Data Recorder GEN3t English User Manual I3995 3 1 en HBM public...
Страница 4: ...4 I3995 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3t I3995 3 1 en HBM public 51...
Страница 1122: ...1122 I3995 3 1 en HBM public...
Страница 1123: ......