Model S4000CH
60
Table 18: ARGC Cable Lengths
* Screw terminals only
Max. distance for analog output (650 Ohms max):
AWG
FEET
METERS
14 9000 2740
16
5200
1585
18 3800 1160
20
2400
730
Table 19: Analog Output Cable Lengths
NOTE:
The analog circuit can tolerate a negative output up to -5 volts.
7.3.5 Remote Sensor Cable Lengths
The remote sensor can have up to 1.5 Ohms in each wire. The wires must be of equal
length and size.
NOTE:
The power wiring length is reduced by 10% due to loss in sensor wiring.
AWG
Ohm/1000 ft
Feet
Meters
12
1.588
1007
307
14
2.525
633
193
16
4.016
398
121
18
6.385
250
76
20
10.15
157
48
Table 20: Sensor Cable Lengths
European Union (EU) Approved Applications:
PSU noise and ripple voltage 1.0 Vpp
max. The customer supplied PSU must comply with IEC 1010-1, limiting current to 8 A
under fault conditions, in order to comply with CE Marking requirements.
7.3.6 Environmental
Specifications
*
With remote junction box 10252-3 and 11159-2 sensor.
Operating Temperature
Range
CSA
FM
ATEX/IECEx
Electronics
Division Classification
-40°F to +167°F (-40°C to +75°C)
-40°F to +392°F (-40°C to +200°C)*
Zone Classification
-40°F to +158°F (-40°C to +70°C)
-40°F to +167°F
(-40°C to +75°C)
-40°F to +158°F
(-40°C to +70°C)
Storage Temperature Range
-58°F to +185°F (-50°C to +85°C)
Operating Humidity Range
0% to 95% RH, non-condensing
Atmosphere
Will not operate in <5% Oxygen. Oxygen enriched reading may be slightly higher