Model S4000CH
55
Address
(hex)
Parameter
Function
Data Type
Data range
Access
5B
Structure time
Low
Hi byte – min, low
byte – sec for event
log entries
Numeric value
0 – 65535
Calibrate
5C
Calibrate code
Cal
Numeric value
0
Calibrate
5D
Reserved
Reserved
Numeric value
0
5E Calibrate
Count
Calibrate
Count
Numeric value
0 – 65535
Calibrate
5F
Reset Event
Counters
Reset Event
Counters
Numeric value
1
Reset
6.8.43 User Information (60h to 6Fh)
There is a section in memory that allows the user to store information. This is useful if the
physical location or other user identification is required. The only restriction on the
information is it must be Modbus compatible. Only one word can be written per command.
There are a total of 16 words for the user.
6.8.44 Total Receive Errors (70h)
A read indicates the total Modbus Comm. Hardware Receive Errors that occurred in the
slave device. The maximum count is 65535 and then the counter will rollover to zero and
begin counting again. The total errors are an accumulation of the individual communication
errors listed below.
6.8.45 Bus Activity Rate % (71h)
A read indicates the Bus Activity Rate in percent of this Slave’s addressed node versus
other addressed nodes. Range of this value is in hex (0-64), which translates to decimal (0-
100%).
6.8.46 Function Code Errors (72h)
A read indicates the number of Function Code Errors that occurred in the slave device. The
maximum count is 65535 and then the counter will rollover to zero and begin counting again.
6.8.47 Starting Address Errors (73h)
The counter is incremented for each address that does not equal the device address. A
read indicates the number of Starting Address Errors that occurred in the slave device. The
maximum count is 65535 and then the counter will rollover to zero and begin counting again.
6.8.48 Number of Register Errors (74h)
A read indicates the Number of Register Errors that occurred in the slave device. The
maximum count is 65535 and then the counter will rollover to zero and begin counting again.