Model S4000CH
52
Calibration
The time a successful zero and calibration occurs is saved in the event log. A counter is
incremented for each successful calibration. If the unit fails Zero or Cal the event is not
recorded. If the calibration is aborted the event is also not recorded. Note the Zero and
Calibration are one operation in the S4000CH. There are a total of 10 event time stamps
stored.
Maintenance
The time a Check Calibration occurs is saved in the maintenance event log. Each
successful calibration increments the maintenance counter. There are a total of 10 event
time stamps stored.
Setting Clock
Please see table below.
Event Logging Registry Table
Address
(hex)
Parameter
Function
Data Type
Data range
Access
30
Seconds Time Hi
Seconds Time Hi
Numeric value
0 – 65535
Timer Sec
31
Seconds Time
Low
Seconds Time
Numeric value
0 – 65535
Timer sec
32
Real Time Clock
Year, Month
Read/Set year and
month of RTC
2 Numeric Values
0-99 year, 1 –
12 month
Timer Struct
33
Real Time Clock
Day, Hour
Read/Set day and
hour of RTC
2 Numeric Values
1 – 31 day, 0
– 23 hour
34
Real Time Clock
Minute, Second
Read/Set minutes
and seconds of
RTC
2 Numeric Values
0 – 59 minute,
0 – 59 second
Timer Struct
35
PowerCycleFlag
Read Power Cycle
Flag.
Numeric Value
1 – Time Not
Reset; 0 –
Time Was
Reset
Flag
36
Event Index
Event index of
Logged Event
Numeric value
0 - 9
Index
37
Warn
Seconds Time Hi
Seconds Time Hi for
warning event log
entries
Numeric value
0 – 65535
Warn
38 Seconds
Time
Low
Seconds Time Low
for warning event
log entries
Numeric value
0 – 65535
Warn
39
Structure time Hi
Hi byte – year, low
byte – month for
warning event log
entries
Numeric value
0 – 65535
Warn