Model S4000CH
44
6.8.8 Mode & Error (07h)
See Mode (02)
Calibration Complete
8000
Spanning
4000
Zero Complete, Waiting for Gas
2000
Zeroing
1000
Calibration Mode
0800
Calibration Check Mode
0400
Run Mode
0200
Startup Mode
Alarm
Warn
Error (any error)
NA
NA
ARGC
Internal
Switch
0100
0080
0040
0020
0010
0008
0004
0002
0001
6.8.9 Error & Sensor Life (08h)
6.8.9.1 Upper 8 Bits for Sensor Errors (Hexadecimal)
Setup error
80
Cal Check error
40
EEPROM Check sum error
20
Flash Check sum
10
Sensor
08
Calibration error
04
Low supply
02
NA
01
6.8.9.2 Lower 8 Bits for Sensor Life
Sensor Life
0-100%
6.8.10 Display (0x09h & 0x0Ah)
The display is also present on the Modbus. It is at address 0x09 and 0x0A. The first
address (0x09) contains the LED, the decimal point location, and the Most Significant Digit
(MSD). The upper word represents the LED and decimal point. They are defined as shown
below. The lower word is the ASCII value for the MSD. The second address (0x0A)
represents the middle digit (MID) and the Least Significant Digit (LSD) in ASCII. The upper
word represents the MID and the lower word represents the LSD.
DP_LSD
0x01
DP_MID
0x02
DP_MSD 0x04
WRN_LED 0x08
ALM_LED 0x10