6.8.37 Overrun errors (28h)
A read indicates the number of hardware UART Overrun Errors that occurred in the slave
device. The maximum count is 65535 and then the counter will rollover to zero and begin
counting again.
6.8.38 Framing Errors (29h)
A read indicates the number of hardware UART Framing Errors that occurred in the slave
device. The maximum count is 65535 and then the counter will rollover to zero and begin
counting again.
6.8.39 Total Software CH1 Errors (2Ah)
A read indicates the number of address or data errors that occurred in the slave device. The
maximum count is 65535 and then the counter will rollover to zero and begin counting again.
6.8.40 New Sensor Calibration (2Bh)
The new sensor calibration command performs a calibration only when the ARGC is
enabled, otherwise it returns an exception. It also tells the ARGC function this is a new
sensor and the data will be saved. Stored data may later be used to verify proper calibration.
NOTE:
All new installations should perform this calibration.
6.8.41 Clear Hardware Errors (2Ch)
For clearing parity, framing and overrun errors.
6.8.42 Clear Communication Errors (2Dh)
6.8.42.1 Event Logging
Faults
Whenever the fault word changes the time will be recorded.
The time of the fault will be saved.
The quantity of faults will be saved in a counter.
When the fault is removed, it is not saved and the counter is not incremented.
A fault event is logged for every 30 seconds recorded.
There are a total of 10 event time stamps stored.
Warning
The time the gas level reaches the Warning level is recorded. Each time this happens a
counter is incremented. The end of the event is when the gas goes below 5%. The
counter is also saved. There are a total of 10 event time stamps stored.
Alarm
The time the gas level reaches the Alarm level is recorded. Each time this happens a
counter is incremented. The end of the event is when the gas goes below 5%. The
counter is also saved. There are a total of 10 event time stamps stored.