Chapter 8 - Replacement Procedures
8-25
6.) Loosen and remove the four screws on the monitor securing bracket (the metal cover was removed
in step 4 to provide easy access) - refer
7.)
Using two people
, lift the monitor upwards, away from the control console.
Note:
While pulling the monitor upwards, swiveling it slightly to the
right
and
left
will ease removal
NOTE:
When removing the monitor, make sure that the Teflon washer remains in place on top
of the monitor mounting base.
8-3-2-4
Vivid™ 4 17" Monitor Installation
Note:
Ensure that the Teflon washer is in place before you return the monitor to the control
console
.
1)
Using two people
, lift the monitor into position on the control console.
2) Fasten the monitor securing bracket in position using the four screws previously removed.
3.) Reconnect the VGA cable and AC power cable to the receptacle cable inside the service
compartment (refer to
4.) Return the six screws and fasten the service cover into position.
5.) Replace the Color printer (if previously removed) from the side compartment.
6.) Reconnect the main AC power cable from the wall outlet.
7.) Turn ON the Vivid™ 4 main circuit breaker, located at the rear of the machine (lower end).
8) Check that the monitor can turn to the
left
and
right
to a limited degree. Also check that it can be
tilted
upwards
and
downwards
, and that it holds its set position.
Figure 8-23 Monitor Securing Bracket