Chapter 8 - Replacement Procedures
8-131
8-8-4-4
Panasonic VCR Installation Procedure
1.) Raise the control console to
maximum
height.
2.) Place the VCR on the Vivid™ 4 ultrasound unit’s top surface (
left
side, below the control console, as
below), allowing sufficient room to access the cables at the rear of the VCR.
Figure 8-125 Panasonic Video Cassette Recorder (VCR) Installed in the Vivid™ 4 Ultrasound Unit
Figure 8-124 Panasonic VCR Switches Setting
Pull Front Switch Lock
Video Cassette
Recorder (VCR)