GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
7-126
Section 7-5 - Common Service Interface
iLinq IP Restore Procedure:
When required (for example, after software re-installation), you can perform Restore of iLinq IP settings
from the hard disk, as follows:
1) From the System Configuration window (under the
Technical Support
above) trackball to
iLinq Settings: Restore
and press
Select.
A blank screen appears; wait for a while until the
phone and modem options
message is displayed
then T
RACKBALL
to
OK
and press
Select
.
2) Wait while the restore process is running - this may take a while.
When complete, a message is displayed prompting for system shutdown.
3.) Perform a full system shutdown, then re-start the Vivid™ 4 system.