GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-133
5.) Place the VCR in its correct final position.
6.) Secure the VCR by fastening the belt around it, or together with the B/W Printer (if present).
7.) If more secure fastening is desired, install the optional shelf (P/N H45011CN), which can be ordered
separately if the Color Printer option is not installed
.
8.) Re-install the Connectors Panels Cover as described in the
Connector Panels Cover Installation
9) Reconnect the mains AC power cable from the wall outlet.
10) Turn ON the main circuit breaker located on rear of the Vivid™ 4.
11) Turn ON the Vivid™ 4 by pressing the ON/OFF button on the Control Console.
12.)Under System Configuration, VCR/ECG Tab, select the Panasonic VCR option, as shown in
below. For more information, see .
Note: Panasonic VCR - Reset Counter:
If the Reset Counter button is activated while the VCR is in the
Stop
state, the counter
display is reset to
zero
(0:00:00).
If activated in either the
Pause
or
Record
state, the display continues running (counter does
not zero).
Figure 8-127 Panasonic VCR Option