GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
8-118
Section 8-8 - Peripherals
8-8-2
Mitsubishi VCR Replacement Procedure
NOTE:
This section describes the installation and removal procedure of the following MITSUBISHI VCR models:
•
HS-MD3000E (PAL) - P/N: 066E9513
•
HS-MD3000U (NTSC) - P/N: 066E9514
The VCR will not operate without initial setup and adjustments, as described below.
Figure 8-105 Mitsubishi Video Cassette Recorder (VCR)
8-8-2-1
Tools
No special tools needed.
8-8-2-2
Manpower
One person - 30 mins.
8-8-2-3
Preparations
1) Shut down the Vivid™ 4 ultrasound unit, as described in
.
2.) Remove the Connectors Panels Cover as described in the
Connector Panels Cover Removal
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.