Chapter 8 - Replacement Procedures
8-117
8-8-1-4-1
Setting the Switches on the Paper Tray
1.) Press the OPEN button in the front of the B/W Video Printer.
2.) Ensure that the switches on the Paper Tray are set as shown in
3.) Load the Paper Roll as described in the Printer’s Operating Instructions.
8-8-1-5
B&W Printer Removal Procedure
1.) Remove the Connectors Panels Cover as described in the
Connector Panels Cover Removal
2.) Release the B/W printer by undoing the fastening belt secured around it (and around the VCR, if
present).
3.) Disconnect the Remote cable, the B/W Video IN cable, and Power cable from rear of the printer
(refer to
4.) Remove the B/W Video Printer from the
right
side of the Vivid™ 4 ultrasound unit (taking it out from
below the control console and from beside the B/W printer, if present) - refer to
Figure 8-104 B/W Video Printer, Slide Switches Location and Settings on the Paper Tray
Slide Switch
Settings on the
Paper Tray
OPEN Tray button