Section 7-3 - Performing Front End (FE) Diagnostics
Figure 7-15 F.B. Tab - FE Diagnostics Dialog Box (RFT Configuration)
3) Trackball to the following buttons (in the order shown) and press
Select
:
a.)
Transmit Test:
Checks the FE pulsers via the probe elements, as described in the
b.)
Receive Test:
Tests each amplifier output against the predefined output level, as described
in the
c.)
TR Switch Test:
Checks the transmit switches, as described in the
d.)
FLA Jumper Test:
Checks the jumpers that connect the side elements to the center elements,
as described in the
e.)
Pencil Probe Test:
Checks the transmitting channels associated relays to the pencil probe
port, as described in the
f.)
Port Test:
Checks general logic control and ports logic, as described in
g.)
RAM Test:
Checks the RAM that stores the delays for the transmitted ultrasound signals to
the form focal beam, as described in the
Data Flow Map
Test Description Area
Test Area
Comment Area