Section 7-4 - Performing Back End Diagnostics on the System
7-4-6
Media Driver Diagnostic Test
The media driver diagnostic test enables you to check the functionality of the MOD, if installed, as well
as the functionality of the CD Writer, if installed.
7-4-6-1
MOD Diagnostic Test
WARNING: Do not run an MOD Diagnostic Test on MO media that is being used by the customer
as a back-up device, since this test may cause unrecoverable damage to any backed-up patient
information on the MO media.
1) Insert a new MO Media Sony EDM-2300B (2.3G) into the MOD.
1) Access the
HW Diagnostics
menu, as described in the
Accessing the Back End Diagnostic Options
2) Trackball to the
Media Driver
button and press
Select
. The
Util_MO_CDR_App
dialog box is
displayed, as shown below:
Figure 7-52 Util_MO_CDR_App Dialog Box
3) Trackball to the
MO
radio button and press
Select
.
4) Trackball to the
IsMediaAccess
button an press
Select
. The message
Media in drive G
is
accessible is displayed.
5) Trackball to the
Eject
button and press
Select
. The message
Media is outside
is displayed. Verify
that the media has actually been ejected.
6) Trackball to the
Load
button and press
Select
. The message
Media is inside
is displayed. Verify
that the media is actually inside.
7) Trackball to the
Format
button and press
Select
. The message
Media in drive G was formatted
successfully
is displayed.
8) Trackball to the
GetVolInfo
button and press
Select
. The message
Media in drive G has
following properties: Media Label = MY_LABEL, Media File-System Name = FAT
is displayed.
9) Trackball to the
GetBtPSector
button and press
Select
.
The message The Bytes Per Sector
value of media in drive G is 512
is displayed.
10) Trackball to the
SetVolLabel1
button and press
Select
.The message
Media Label set to
TEST_LABEL1
is displayed.