Chapter 8 - Replacement Procedures
8-23
5.) Unscrew and remove each of the 4 screws that fasten the Monitor Rear Cover in position (two each
side; one
upper
and one
lower
6.) Carefully lift the Monitor Rear Cover
upwards
and
backwards
to remove it from the monitor.
8-3-1-4
Vivid™ 4 17" Monitor Rear Cover Installation
1.) Make sure the Vivid™ 4 control console is lowered to its
minimum
height.
2.) Carefully lift the replacement Monitor Rear Cover into position on the monitor.
3.) Return and fasten each of the 4 screws (previously removed) to secure the Monitor Rear Cover
firmly (two each side; one
upper
and one
lower
4.) Return the two screw covers (previously removed) to the top of the Monitor Rear Cover (one each
side) and place in the
closed
position - refer to
.
5.) Reconnect the AC power cable to the mains wall outlet.
6.) Turn ON the Vivid™ 4 main circuit breaker, located at the rear of the machine (lower end).