background image

Technical 
Publication

Part Number 2392751-100
Revision 3           

GE Medical Systems
Vivid™ 4 Service Manual

Copyright

©

 2006 by GE Medical Systems

Содержание Vivid 4 BT03 RFI

Страница 1: ...Technical Publication Part Number 2392751 100 Revision 3 GE Medical Systems Vivid 4 Service Manual Copyright 2006 by GE Medical Systems...

Страница 2: ...UE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS T CONSULT ET COMPRIS LE NON RESPECT DE CET AVERTISSEM...

Страница 3: ...NUAIS NOUTRO IDIOMA DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVI OS DE TRADU O N O TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSIST NCIA T CNICA O N O CUMPRIMENT...

Страница 4: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL iii...

Страница 5: ...tors Other connections between pieces of electrical equipment calibrations and testing shall be performed by qualified GE Medical Systems personnel In performing all electrical work on these products...

Страница 6: ...ted in any form in whole or in part without prior written permission of GE Medical Systems GE Medical Systems may revise this publication from time to time without written notice TRADEMARKS All produc...

Страница 7: ...ge N A Chapter 2 Pre Installation pages 2 1 to 2 12 3 Chapter 7 Diagnostics Troubleshooting pages 7 1 to 7 132 3 Important Precautions pages i to iv 3 Chapter 3 Installation pages 3 1 to 3 80 3 Chapte...

Страница 8: ...s 1 4 Important Conventions 1 5 Conventions Used in this Manual 1 5 Model Designations 1 5 Icons 1 5 Safety Precaution Messages 1 5 Standard Hazard Icons 1 6 Safety Considerations 1 7 Introduction 1 7...

Страница 9: ...IRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL viii Electrostatic Discharge ESD Prevention 1 24 Standards Used 1 25 Lockout Tagout Requirements For USA Only 1 25 Customer Assistance 1 26 Conta...

Страница 10: ...ite Circuit Breaker 2 3 Site Power Outlets 2 4 Mains Power Plug 2 4 Power Stability Requirements 2 4 EMI Limitations 2 5 Probe Environmental Requirements 2 6 Facility Needs 2 7 Purchaser Responsibilit...

Страница 11: ...paring for Installation 3 13 Confirming Customer Order 3 13 Verifying the Shipping Crate Contents 3 13 External Cables 3 13 Accessories 3 14 Component Inspection 3 14 Damage Inspection Checklist 3 14...

Страница 12: ...s Tab 3 40 System Options Tab 3 42 Printers Tab 3 44 VCR ECG Tab 3 45 Technical Support Tab 3 47 Technical Support Tab RFI Configuration 3 47 Technical Support Tab RFT Configuration 3 49 Technical Sup...

Страница 13: ...pping Crate and Packaging Materials 3 71 Cardboard Shipping Carton and Packaging Materials 3 72 Packing the Unit into the Wooden Shipping Crate 3 73 Assembling the Wooden Shipping Crate 3 74 Packing t...

Страница 14: ...rinter Tests 4 5 Mechanical Functions 4 6 Console Movement Test 4 6 Brakes Function Test 4 6 Back End Processor Tests 4 7 Internal ECG Test 4 7 Phono Test optional 4 7 Magneto Optical Drive MOD Test 4...

Страница 15: ...739L Probe Image Quality Tests 4 21 2D Noise Uniformity Test 4 21 2D Performance Test 4 21 CFM Stationary Noise Test 4 22 Probe Operation Check Alignment Sensitivity 4 22 CFM Operation Tests 4 23 PW O...

Страница 16: ...I configuration only 5 12 Transmit Signal Path 5 12 Received Signal Path 5 12 Front Board Assembly FB 5 16 General Description 5 16 TR4 Board Description 5 17 MUX Board 5 18 General Description 5 18 B...

Страница 17: ...5 38 Floppy Drive 5 39 Hard Disk 5 39 Magneto Optical Drive MOD 5 40 CD Read Write CDRW 5 40 ECG Module 5 40 Modem 5 41 PC VIC Assembly 5 42 General Description 5 42 PAL 5 43 NTSC 5 43 External Periph...

Страница 18: ...vice Platform 5 52 Introduction 5 52 iLinq Interactive Platform Features 5 52 Web Server Browser 5 52 Connectivity 5 52 Configuration 5 52 Contact GE 5 52 Interactive Application 5 52 Global Service U...

Страница 19: ...rt IMP LEDs 6 5 Front End Controller FEC LEDs 6 6 RF and Tissue Processor RFT 6 6 Beamformer BF 6 6 Channels Multiplexer MUX 6 7 Front Board Assembly FB 6 7 Back End Power Supply Voltages 6 8 VIC Vide...

Страница 20: ...c Tests for RFT Configuration 7 15 RFT Diagnostic Tests for RFT Configuration 7 17 Front End Controller FEC Diagnostic Tests for RFT Configuration 7 19 Beamformer BF Diagnostic Tests 7 21 Front Board...

Страница 21: ...ry 7 76 General Configuration 7 76 Keyboard Diagnostic Test 7 77 Media Driver Diagnostic Test 7 78 MOD Diagnostic Test 7 78 CD Writer Diagnostic Test 7 79 Computer Diagnostic Test 7 80 Benchmark Testi...

Страница 22: ...7 106 Monitor Test 7 106 Front End Diagnostics 7 107 Image Quality Page 7 111 Calibration Page 7 111 Configuration Page 7 112 Utilities Page 7 113 Windows 2000 Utility 7 113 Scanner Utilities 7 115 P...

Страница 23: ...re 8 6 Rear Cover Replacement Procedures 8 7 Tools 8 7 Preparations 8 7 Rear Cover Removal Procedure 8 7 Rear Cover Installation Procedure 8 7 Connector Panels Cover Replacement Procedures 8 8 Tools 8...

Страница 24: ...r Removal Procedure 8 19 Control Console Lower Rear Cover Installation Procedure 8 19 Front Handle Replacement Procedure 8 20 Tools 8 20 Preparation 8 20 Front Handle Removal Procedure 8 20 Front Hand...

Страница 25: ...eaker Replacement Procedure 8 34 Tools 8 34 Preparation 8 34 Speaker Removal Procedure 8 34 Speaker Installation Procedure 8 34 Front End Parts Replacement 8 35 Front End Boards Replacement Procedure...

Страница 26: ...ure 8 54 BEP2 Side Cover Installation Procedure 8 54 BEP2 Front Cover Removal Procedure 8 54 BEP2 Front Cover Installation Procedure 8 56 BEP2 Octopus Card Holder Removal Procedure 8 56 BEP2 Octopus C...

Страница 27: ...ols 8 66 Preparation 8 66 CDRW Drive Removal Procedure 8 66 CDRW Drive Installation Procedure 8 67 MO Drive Replacement Procedure 8 68 Tools 8 68 Preparation 8 68 MO Drive Removal Procedure 8 68 MO Dr...

Страница 28: ...cedure 8 88 AC BEP or FE Cable Installation Procedure 8 88 Gas Spring Cable Replacement Procedure 8 89 Preparation 8 89 Gas Spring Cable Removal Procedure 8 89 Gas Spring Cable Installation Procedure...

Страница 29: ...parations 8 118 Mitsubishi VCR Installation Procedure 8 119 Mitsubishi VCR Setup Procedure 8 122 Mitsubishi VCR Removal Procedure 8 125 Sony VCR Replacement Procedure 8 126 Tools 8 126 Manpower 8 126...

Страница 30: ...MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL xxix HP 6540 3 USB Deskjet Color Printer Installation Procedure 8 140 HP 6540 3 USB Deskjet Color Printer Removal Procedure 8 1...

Страница 31: ...Renewal Parts Overview 9 1 Purpose of Chapter 9 9 1 List of Abbreviations 9 2 Renewal Parts Lists and Diagrams 9 3 Mechanical Hardware Parts 9 3 AC System Parts 9 9 Front End Parts 9 11 Back End Part...

Страница 32: ...ance 10 6 System Periodic Maintenance 10 7 Preliminary Checks 10 7 Functional Checks 10 8 System Checks 10 8 Peripheral Option Checks 10 9 Input Power Checks 10 9 Mains Cable Inspection 10 9 Cleaning...

Страница 33: ...Lead Source Leakage Lead to Ground 10 24 Definition 10 24 Generic Procedure 10 24 Isolated Patient Lead Source Leakage Lead to Lead 10 25 Isolated Patient Lead Sink Leakage Isolation Test 10 25 Data...

Страница 34: ...Chapter 3 Installation Contains installation procedures and an installation checklist Chapter 4 Functional Checks Contains functional checks that are recommended as part of the installation procedure...

Страница 35: ...Parts Contains a complete list of field replaceable parts for the Vivid 4 ultrasound unit Chapter 10 Periodic Maintenance Provides periodic maintenance procedures for the Vivid 4 ultrasound unit 1 1 3...

Страница 36: ...and below are configured with RFT hardware Table 1 2 Vivid 4 BT03 RFI Models Model Cat No Description BASE Vivid 4 Console Vivid 4 BT03 console 220 240V AC RFI H45011FT An advanced version of the Ultr...

Страница 37: ...und unit and therefore shares much of the same hardware However even though much of the hardware is the same the Vivid 3 cannot be upgraded to Vivid 4 or vice versa The Vivid 4 ultrasound unit provide...

Страница 38: ...ls of safety precautions are found on the equipment and throughout this service manual Different levels of severity are identified by one of the following icons which precede precautionary statements...

Страница 39: ...eral different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedures that should be followed...

Страница 40: ...HAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE DO NOT USE A DAMAGED OR DEFECTIVE PROBE FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN...

Страница 41: ...e shared with other types of equipment Both the system power cable and the power connector must meet international electrical standards Note When replacing UPS batteries ensure that defective batterie...

Страница 42: ...DANGEROUS VOLTAGES CAPABLE OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING EXPLOSION WARNING DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSP...

Страница 43: ...rtinent information regarding the operation of the ultrasound unit 1 4 1 Product Label Locations The following two diagrams indicate the location of some of the labels and icons found on the Vivid 4 u...

Страница 44: ...EDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 11 Figure 1 2 Product Label and Icon Locations Rear 1 Main Label 2 AC Voltage Rating Label 3 GND Label 1...

Страница 45: ...cannot become live in the event of a failure of the basic insulation Rear of the unit and probe connectors Device Listing Certification Labels Laboratory logos or labels that denote conformity with in...

Страница 46: ...the lightning flash with arrowhead in equilateral triangle is used to indicate electric shock hazards Rear of the unit Protective Earth Indicates the protective earth grounding terminal Rear of the un...

Страница 47: ...Label section on page 1 23 Swivel Brake Label section on page 1 23 1 4 3 1 Main Label The main label may be printed in any of the following languages English German French Spanish Portuguese Italian C...

Страница 48: ...ION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 15 German Used in all German language countries Figure 1 5 Main Label German French Used in all French language countries Fig...

Страница 49: ...REVISION 3 VIVID 4 SERVICE MANUAL 1 16 Section 1 4 Product Labels and Icons Spanish Used in all Spanish language countries Figure 1 7 Main Label Spanish Portuguese Used in all Portuguese language cou...

Страница 50: ...SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 17 Figure 1 9 Main Label Portuguese 220 240V Italian Used in all Italian language countries Figure 1 10 Main La...

Страница 51: ...1 100 REVISION 3 VIVID 4 SERVICE MANUAL 1 18 Section 1 4 Product Labels and Icons Chinese Used in all Chinese language countries Figure 1 11 Main Label Chinese Danish Used in all Danish language count...

Страница 52: ...ON 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 19 Dutch Used in all Dutch language countries Figure 1 13 Main Label Dutch Finnish Used in all Finnish language countries Figu...

Страница 53: ...1 100 REVISION 3 VIVID 4 SERVICE MANUAL 1 20 Section 1 4 Product Labels and Icons Greek Used in all Greek language countries Figure 1 15 Main Label Greek Japanese Used in all Japanese language countri...

Страница 54: ...2751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 1 Introduction 1 21 Norwegian Used in all Norwegian language countries Figure 1 17 Main Label Norwegian Russian Used in all Russian language countrie...

Страница 55: ...d in all Swedish language countries Figure 1 19 Main Label Swedish 1 4 3 2 Rating Labels Indicates the ultrasound unit s factory preset input AC voltage as follows AC 100V AC 120V AC 220 240V AC 220 2...

Страница 56: ...es the locked pedal position which locks the front castors and prevents the ultrasound unit from moving The parking label shown below is located on the brake pedal at the front of the unit as shown in...

Страница 57: ...produces unacceptable levels of emission This interference is often referred to as radio frequency or electromagnetic interference RFI EMI and can be radiated through space or conducted over interconn...

Страница 58: ...l is required in order to achieve full EMC performance 1 5 4 Lockout Tagout Requirements For USA Only Follow OSHA Lockout Tagout requirements by ensuring you are in total control of the electrical Mai...

Страница 59: ...ltrasound Service Engineering 4855 W Electric Avenue Milwaukee WI 53219 Customer Answer Center Phone 1 800 437 1171 Phone 1 800 321 7937 Phone 1 800 682 5327 Phone 1 262 524 5698 Fax 1 414 647 4125 La...

Страница 60: ...he installation of a Vivid 4 ultrasound unit Included are descriptions of the electrical and facility requirements that must be met by the purchaser A worksheet is provided at the end of this chapter...

Страница 61: ...if the room lighting is subdued and indirect when a scan is being performed Therefore a combination lighting system dim bright is recommended Keep in mind that lighting controls and dimmers can be a...

Страница 62: ...ease note that image artifacts can occur if at any time within the facility the Ground from the main facility s incoming power source to the Ultrasound unit is only a conduit 2 2 5 1 Vivid 4 Power Req...

Страница 63: ...t all current local state and national electrical codes 2 2 5 5 Mains Power Plug If the unit arrives without a power plug or with the wrong plug contact your GE dealer When necessary the installation...

Страница 64: ...wever there is no guarantee that interference will not occur in a particular installation Note Possible EMI sources should be identified before the unit is installed and should not be on the same line...

Страница 65: ...oose or missing covers or RF gaskets allow radio frequencies to interfere with the ultrasound signals Replace broken RF gaskets If more than 20 or a pair of the fingers on an RF gasket are broken repl...

Страница 66: ...alibration and testing must also be performed by qualified personnel The products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is...

Страница 67: ...the case of a network option An active network outlet in the vicinity of the ultrasound unit A network cable of appropriate length regular Pin to Pin network cable An IT administrator who will assist...

Страница 68: ...rator with clinically useful features for moving images and patient information over a hospital network Examples of DICOM services include the transfer of images to workstations for viewing or transfe...

Страница 69: ...ork administrator must provide the required information which should include the following Vivid 4 Details DICOM network details for the Vivid 4 unit including the host name local port IP address AE t...

Страница 70: ...Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept Vivid 4 SN Type Floor Room REV Comments CONTACT INFORMATION Name Title Phone...

Страница 71: ...from the basic installation Make sure system and probes have been subject to acclimation period Environmental cooling is sufficient Lighting is adjustable to adapt to varying operational conditions of...

Страница 72: ...epared as described in Chapter 2 Pre Installation Once the site has been prepared installation can proceed as described in this chapter Table 3 1 Contents in Chapter 3 Section Description Page Number...

Страница 73: ...troubleshooting 3 2 2 1 System Acclimation Time Following transport the Vivid 4 system may be very cold or hot Allow time for the system to acclimate before being switched ON Acclimation requires 1 ho...

Страница 74: ...connect the unit to a properly grounded power outlet Do NOT use a three prong to two prong adapter as this defeats safety grounding CAUTION Do NOT wear the ESD wrist strap when you work on live circui...

Страница 75: ...ge 3 9 3 3 1 Unpacking the Wooden Shipping Crate The Vivid 4 ultrasound unit is packed in a wooden crate that has four walls left right front and rear the crate base and the top cover Each section has...

Страница 76: ...either the Drop or Tilt indicators have turned red showing failure please inform the GE Medical Systems sales representative immediately In addition mark on the shipping consignment note or packing sl...

Страница 77: ...the probes and peripheral options items D and E in Figure 3 3 below from the wooden crate 5 Remove the console control support back foam item 11 in Figure 3 3 above 6 Remove the keyboard support foam...

Страница 78: ...ntrol console monitor by pressing the release grip located under the unit s front handle 15 Remove the two cartons containing the external cables and accessories located under the control console and...

Страница 79: ...wheels from swiveling as described in the Swivel Brake Label section page 1 23 18 Pull the unit backwards from the platform base onto the ramp and roll it down to the floor 19 Press the brake pedal lo...

Страница 80: ...on the Shock watch and Tilt watch labels respectively for evidence of accidental shock or tilting of the crate during transit refer to Figure 3 2 on page 3 5 NOTE If the crate is damaged or if either...

Страница 81: ...ews item 16 so that the cardboard carton item 12 is free of the wooden platform item 4 and the narrow flap is free 4 Open the narrow side flap of the carton remove the ramp item 5 and attach it to the...

Страница 82: ...SERVICE MANUAL Chapter 3 Installation 3 11 8 Pull out and remove the monitor support item 10 in Figure 3 9 below from under the base of the monitor Figure 3 8 Front and Side View of the Unit in the Sh...

Страница 83: ...se cartons contain the external cables and accessories 12 Lower the control console to its lowest position 13 Remove the manuals item 4 in Figure 3 8 on page 3 11 from the compartment on the left side...

Страница 84: ...tallation 3 4 2 1 External Cables The ECG cable item number 10 or 11 is supplied according to installation location Table 3 4 Vivid 4 External Cables Item Number P N Description Quantity 1 2269430 PWR...

Страница 85: ...erial number printed on the main label into the system installation details card as described in System Installation Details on page 3 78 3 4 3 1 Damage Inspection Checklist Visually inspect the conte...

Страница 86: ...and peripheral fans are operating and clean 7 BE Rear Panel Check the BE rear panel connectors for bent pins loose connections and loose or missing hardware Screw all the cable connectors tightly to t...

Страница 87: ...de of the front panel 4 Control Panel Contains the alphanumeric keyboard and the buttons used to operate the ultrasound unit 5 Front Handle 6 Raise Lower the Control Console Up Down Swivel Handle Loca...

Страница 88: ...I O module that enables keyboard commands to be operated by foot 11 Gel Holders Situated on either side of the control console and on the top surface on either side of the monitor 12 Color Printer or...

Страница 89: ...Unit The following figure shows the Vivid 4 ultrasound unit components that are visible from the rear of the ultrasound unit 1 Monitor 2 Gel Wells 3 Monitor Connection Panel 4 Left Rear Panel 5 Right...

Страница 90: ...stem WARNING CONNECTING A Vivid 4 SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY DESTROY THE SCANNER 3 4 4 2 Confirming System Voltage Configuration 1 Turn ON the system 2 In regular 2D Scanning...

Страница 91: ...100 REVISION 3 VIVID 4 SERVICE MANUAL 3 20 Section 3 4 Preparing for Installation 5 Click OK Note If the voltage is not set correctly contact an authorized GE Service Representative Figure 3 14 System...

Страница 92: ...ed 3 4 5 1 Video Format Confirmation 1 Turn ON the system and the check video transmission signal is set correctly 2 Press Config 3 From the System Configuration dialog box select the VCR ECG tab and...

Страница 93: ...y images Check the grounding of the power cord and power outlet Be aware of RF sources Keep the unit at least 5m 16 4 ft away from other EMI sources Special shielding may be required to eliminate inte...

Страница 94: ...CAUTION The Vivid 4 weighs 170 kg 375 lbs or more depending on installed peripherals when ready for use Care must be used when moving it or replacing its parts Failure to follow the precautions liste...

Страница 95: ...formation for each peripheral device refer to the manufacturer s manual On board peripherals must be connected to one of the two available auxiliary power supplies on the right rear panel The total lo...

Страница 96: ...ntry voltage 2 AUXILIARY AC OUTLET For use with external peripherals Voltages are set according to local country voltage If additional auxiliary outlets are required use the special cable provided by...

Страница 97: ...Y C Video Out 4 pin connector for output to an S VHS VCR 3 S VIDEO IN Y C Video In 4 pin connector for input from an S VHS VCR 4 VIDEO OUT B W BNC connector for composite B W video output to a hard c...

Страница 98: ...re 8 101 on page 115 3 After connecting the remaining peripherals and switching the system on configure the printer settings as described in the Printers Tab section on page 3 44 For more details abou...

Страница 99: ...4 unit WARNING Whenever moving the Vivid 4 scanner the DeskJet Color Printer must be disconnected from the scanner DO NOT attempt to move the two units simultaneously without first disconnecting them...

Страница 100: ...Gently push the connector in as far as possible 5 Rotate the locking latch 90 degrees clockwise to lock the connector into place vertical NOTE It is not necessary to turn OFF power to connect or disco...

Страница 101: ...erent cable parts The main part trunk is a single cable connecting to the unit at one end and providing a cable splitter device at the other end The splitter contains five receptacles only three of wh...

Страница 102: ...rds see Table 3 16 Figure 3 19 ECG Cable and Electrode Placement Table 3 16 ECG Cable Types Description Part No Full ECG Cable Kit AHA Americas Japan 2256477 Black wire LA 2269982 3 White wire RA 2269...

Страница 103: ...e wall outlet plug should be connected to the fixed power socket that has a protective grounding conductor Never use an adapter or converter to connect with a power source plug for example a three pro...

Страница 104: ...g the power cord connector into the AC input socket and secure it in place using the attached clip see Figure 3 20 4 Plug the other end of the power cord to the AC wall outlet Allow sufficient slack s...

Страница 105: ...obe and application that were last used before the system was shut down If the probe has been removed since the system was last used the currently connected probes and their available applications are...

Страница 106: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 3 Installation 3 35 Figure 3 21 Shut Down Options Screen...

Страница 107: ...ght They can be adjusted using the Contrast and Brightness buttons on the front part of the display monitor All display monitor controls other than the contrast and brightness controls are factory adj...

Страница 108: ...area set the date and time as follows Trackball to the Set button and press Select Use the alphanumeric keyboard to select the current date Select the format of the date and time display for example D...

Страница 109: ...gure 3 24 Archive Tab Note The Archive tab will display even if the Archive Package option is not installed 3 Select Display patients gender on screen to display the patient s gender in the patient in...

Страница 110: ...to which data is stored during backup from the Backup Device Is dropdown list 10 Select Eject MO Disk on Shutdown to have the system automatically eject the backup media when the unit is shut down 11...

Страница 111: ...name is highlighted Trackball to the Change button and press Select The Font dialog box is displayed Trackball to the required font color style and size and press Select for each choice Trackball to...

Страница 112: ...layed it will appear in the newly selected position until such time as this setting is changed 5 Activate the Duplicate Bodymark on Switch Window and or the Bodymark Menu on Freeze checkboxes as requi...

Страница 113: ...xt to the name of the option being installed and use the alphanumeric keyboard to enter the password supplied by GE Note If an incorrect password is entered if the option has already been correctly in...

Страница 114: ...43 Note If an option was not initially purchased or installed it may be acquired and or installed later During option installation type the relevant password in the appropriate Option window see Figu...

Страница 115: ...with the Vivid 4 system Other generic printers are not approved and GE Medical Systems will not guarantee their correct operation on this system 3 In the Currently Installed Printers area trackball to...

Страница 116: ...ions area complete the following as required Select the video system from the Video Mode dropdown list for example PAL or NTSC Select the SmartVCR option to enable Measurement and Analysis measurement...

Страница 117: ...d define the default length of a cineloop when there is no ECG trace Select the Analyze only Systole checkbox to place markers only on the systole portion for synchronization and analysis purposes Sel...

Страница 118: ...Tab RFI Configuration Monitoring should always be enabled this provides constant monitoring of critical parameters to ensure safe operation of the system Watchdog normally should not be enabled When...

Страница 119: ...onfiguration Table 3 17 Recommended Vivid 4 Hardware Configurations RFI Models Item Vivid 4 BT03 ECG Second Re Image Port No board Front Board V7 3 TR V4 Ex Keyboard Revision 4 MUX Board V2 Sound Boar...

Страница 120: ...hdog normally should not be enabled When enabled if there is a system lock up caused by the Back End this will turn OFF power at the Front End MOD Fan should be enabled only on BEPs based on Pentium I...

Страница 121: ...ration Table 3 18 Recommended Vivid 4 Hardware Configurations RFT Models Item Vivid 4 BT03 ECG Second Re Image Port Revision E8 F8 Front Board V7 3 TR V4 Ex Keyboard Revision 4 MUX Board V2 Sound Boar...

Страница 122: ...r 3 Installation 3 51 3 6 11 Technical Support History Tab The Technical Support History is where you would record any historical technical support information that may be relevant to a problem or any...

Страница 123: ...station or a DICOM service using a TCP IP connection 3 7 2 Physical Connection 3 7 2 1 Ethernet Switch Connections An Ethernet Switch P N 066E0741 is used to connect the Vivid 4 system to an EchoPAC P...

Страница 124: ...Workstation Figure 3 34 Hospital Network Connections See the EchoPAC PC Service Manual Part Number EP091298 for details about the EchoPAC PC and the required set up procedures EchoPAC PC Vivid7 Do no...

Страница 125: ...s to the Hospital Information System HIS through HL7 protocol Users requiring this option should contact the GE representative 1 Press Config on the alphanumeric keyboard The System Configuration wind...

Страница 126: ...L Chapter 3 Installation 3 55 4 Enable the DICOM printer option as required under the Services sub tab see Figure 3 36 Figure 3 36 Setting up DICOM Print 5 Enable any other remote server as required f...

Страница 127: ...AL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 3 56 Section 3 7 Connectivity Setup Figure 3 37 Setting up Remote Server for MPEGvue 6 When done trackball to the OK button and press...

Страница 128: ...d 2 Trackball to the Connectivity tab and press Select The Connectivity tab is displayed with the Views subsidiary tab selected by default as shown below Figure 3 38 Connectivity Tab 3 Trackball to th...

Страница 129: ...omputer Name field shows the serial number of your Vivid 4 system 6 Enter the network information in the appropriate fields Leave the User Name as vivid3 Note The Default Gateway option is not always...

Страница 130: ...window is displayed 3 Trackball to the Connectivity tab and press Select The Connectivity tab is displayed with the Views subsidiary tab selected by default as shown in Figure 3 38 on page 3 57 4 Trac...

Страница 131: ...rage Server line in the Services table The Services tab is expanded as shown below Figure 3 43 Connectivity Tab Services for DICOM Server Expanded 10 Enter the AE Title and the Port No of the DICOM se...

Страница 132: ...er 3 Installation 3 61 Figure 3 44 Connectivity Tab Settings for DICOM Print Expanded 12 Under Settings select the required Format Film Orientation etc see Figure 3 44 above 13 Proceed to the next sec...

Страница 133: ...taflow configuration such as Int HD MOD 525 DICOM images are stored both on the internal hard drive and on media MOD The name DICOM is shown in parenthesis signifying that on the MOD the images are in...

Страница 134: ...shed a red cross is displayed The properties of the DICOM storage server configuration are displayed on the right side of the tab for reference 6 If the connection was successfully established trackba...

Страница 135: ...l Archive Int HD DICOM Server option from the drop down list at the bottom of the window If the system will not accept the dataflow or displays an error message there is an error in dataflow communica...

Страница 136: ...reject any images that have a future date Ensure that the clock on the Vivid 4 ultrasound unit lags behind that of the EchoPAC system by a minute or two 1 Press Config on the alphanumeric keyboard Th...

Страница 137: ...ver from the Name dropdown list The Services tab is updated as shown below Figure 3 48 Services Tab EchoPAC 8 Enter the IP address of the EchoPAC server in the IP address field 9 Select the Remote Arc...

Страница 138: ...hive in the Placement of Report Files field OR If Image Vault Server is required make sure that the Image Vault Server checkbox is selected as shown in Figure 3 50 below then proceed as follows a Ente...

Страница 139: ...mote Storage is displayed 5 Trackball to Remote Servers and press Select then trackball to the Check button and press Select If the connection is established a green checkmark is displayed If the conn...

Страница 140: ...ew services to a dataflow In the event that any changes or additions are required contact the On line Center 3 7 6 2 Validating EchoPAC or Image Vault Operation 1 Press the Archive button on the contr...

Страница 141: ...m is completely shut down Power disconnection before full system shutdown may cause system boot up failure This may require system recovery by booting up the system using the software CD Before storin...

Страница 142: ...an ten degrees to prevent the unit from tipping over Once the destination is reached lock the front wheel brakes 3 8 4 Wooden Shipping Crate and Packaging Materials Verify that the following packing m...

Страница 143: ...other means Table 3 21 Packaging Materials for the Cardboard Shipping Crate Item Description Qty 1 Press on envelope 2 2 Product Label 1 3 Input Pwr Cable 1 4 Platform wooden base 1 5 Ramp 1 6 Antista...

Страница 144: ...al and any other manuals in an antistatic bag item 4 11 Lay the case base on the floor with the two wood strips facing upwards 12 Cover the top side of the base with the antistatic sheet 2 2 m x 3m 7...

Страница 145: ...d appear next to each other in one corner The top cover has also a red mark that should be facing downwards in the corner where the other three marks meet Note Always insert the short leg of a clip in...

Страница 146: ...in place using eight clips item 15 11 Secure the hand lever clip remover tool item 23 to the outside of the wooden crate using a screw 12 Secure the box with two steel strip loops item 16 around the...

Страница 147: ...in an antistatic bag 11 Lay the wooden platform item 4 on the floor Leave sufficient free space around the platform 12 Lay the ramp item 5 on the floor and attach it to the narrow side of the wooden...

Страница 148: ...included 1 Ensure that the ultrasound unit was properly packed that the brake pedal is in the parking position and that the front stopper is locked and prevents the unit from moving on the wooden plat...

Страница 149: ...uals Installation Manuals for the peripheral units must be kept as part of the original system documentation This will ensure that all relevant safety and user information is available during the oper...

Страница 150: ...he installation Specific language versions of the Vivid 4 User Manual may also be available Check with your GE Sales Representative for availability During and after installation supplementary documen...

Страница 151: ...nstallation procedure or as required during servicing and periodic maintenance of the Vivid 4 ultrasound scanner Note Refer to Chapter 7 Diagnostics Troubleshooting for detailed descriptions of some o...

Страница 152: ...Listen to fan spinning sound check filter go to monitoring and verify that all fans are working 2 Connect each of the probes available for the system Verify that the system properly recognizes each o...

Страница 153: ...lider is moving on the Diagnostic panel 8 Perform an automatic lights test as described in the Automatic Lights Test on page 7 75 Verify that all the key LEDs are illuminated one by one 9 Select the S...

Страница 154: ...able and has similar grey levels to that of the reference image Table 4 7 VCR Test Step Task Expected Result s 1 Connect a Sony 9600 VCR to the system restart the system and load an SVHS cassette Veri...

Страница 155: ...image See B W Printer Tests below The image is displayed correctly 5 Connect the monitor to the Color Printer RGB connectors and check the image See Color Printer Tests below The image is displayed co...

Страница 156: ...t you do not apply too much force to move the console and that the console movement is smooth 4 With the front handle still in the fully pressed position push the console to its minimum height Ensure...

Страница 157: ...nd electrode pads 2 Perform the ECG diagnostic test as described in the ECG Phono Diagnostic Test on page 7 72 Verify that a single clear QRS signal is displayed Table 4 14 Phono Test Step Task Expect...

Страница 158: ...y is displayed 8 Select the GetVolInfo button The message Media in drive G has following properties Media Label MY_LABEL Media File System Name FAT is displayed 9 Select the Get BTPsector button The m...

Страница 159: ...abel Media File System Name CDUDF is displayed 8 Press the Get BTPsector button and respond OK to the message The message The Bytes Per Sector value of Media in Drive H is 2048 is displayed 9 Press th...

Страница 160: ...he system remains in Standby mode indicated by the blinking On Off button LED 3 Turn ON the system s circuit breaker Press the On Off Button The system returns quickly to the Power ON state less than...

Страница 161: ...effects 4 4 1 3S Probe Image Quality Tests 4 4 1 1 2D Noise Uniformity Test 4 4 1 2 2D Performance Test Table 4 18 2D Noise Uniformity Test 3S Step Task Expected Result s 1 With a 3S probe in the air...

Страница 162: ...the phantom at position 3 Gain 37 Power 0 dB Depth 20cm Focus Max depth Dynamic Range 65dB 4 Record the maximum depth at which tissue can be differentiated from the noise The depth should be greater t...

Страница 163: ...he Electronic Flow Phantom RMI1425A 2 Place the 3S probe above the vertical nails line position 1 3 Select a Cardiac preset and activate CFM 4 Set the following CFM parameters Gain Max possible withou...

Страница 164: ...t and set the 2D Gain to 40 2 Activate CFM and set the following CFM parameters Gain Max possible without color noise Frame Rate 2 LVR MIN setting CFM Artifacts Test 3 Place the 3S probe on the RMI142...

Страница 165: ...the following 2D parameters 2D Gain 50 2D Dynamic Range 60 2D Reject 10 2 Activate CFM and set the CFM Gain to the maximum possible without color noise 3 Activate M Mode Color M Mode CMM is activated...

Страница 166: ...area is dark and that there are no flash strikes or other artifacts 5 Move the sample volume to different locations along the vessel Ensure that the flow looks clear the noise area is dark and that t...

Страница 167: ...oise area is dark and that there are no flash strikes or other artifacts 5 Move the CW cursor to different locations along the vessel Ensure that the flow looks clear the noise area is dark and that t...

Страница 168: ...hat the Doppler audio and display noise are homogenous and constant 3 Move the cursor angle multiple times from one half of the screen to the other Check that the noise level does not increase after t...

Страница 169: ...parameters Depth 22cm Frame Rate 2 single focus Focus Location 12cm 3 Look for artifacts in the image Table 4 31 CFM Artifacts Test C358 Step Task Expected Result s 1 Use the Electronic Flow Phantom...

Страница 170: ...w looks clear the noise area is dark and that there are no flash strikes or other artifacts 5 Move the sample volume to different locations along the vessel Ensure that the flow looks clear the noise...

Страница 171: ...options 2 Set the following parameters 2D Gain Max Depth 2cm Gain Sliders Max Up to 5mm of white carpet noise should be displayed as shown in the left figure above Ensure that there are no white lines...

Страница 172: ...the Electronic Flow Phantom RMI1425A 2 Place the 739L probe above the vertical nails line position 1 3 Select a Cardiac preset and activate CFM 4 Set the following CFM parameters Gain Max possible wi...

Страница 173: ...eration Tests 739L Step Task Expected Result s 1 Select a Cardiac preset and set the 2D Gain to 40 2 Activate CFM and set the following CFM parameters Gain Max possible without color noise Frame Rate...

Страница 174: ...w looks clear the noise area is dark and that there are no flash strikes or other artifacts 5 Move the sample volume to different locations along the vessel Ensure that the flow looks clear the noise...

Страница 175: ...nter of the screen as shown opposite and write down the Gain value 4 Increase the Gain until background noise is displayed and write down the Gain value Ensure that the Gain difference is not higher t...

Страница 176: ...4 6 show example configurations and should be used for reference only as each site has its own specific settings 4 5 1 Software Configuration Checks Table 4 42 Software Configuration Checklist Step T...

Страница 177: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 4 Functional Checks 4 27 Figure 4 1 System Tab Configuration Example...

Страница 178: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 4 28 Section 4 5 System Turnover Checklist Figure 4 2 System Options Tab Configuration Example...

Страница 179: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 4 Functional Checks 4 29 Figure 4 3 Hospital Info Tab Configuration Example...

Страница 180: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 4 30 Section 4 5 System Turnover Checklist Figure 4 4 VCR ECG Tab Configuration Example...

Страница 181: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 4 Functional Checks 4 31 Figure 4 5 Archive Tab Configuration Example...

Страница 182: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 4 32 Section 4 5 System Turnover Checklist Figure 4 6 Connectivity Local Settings Tab Configuration Example...

Страница 183: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 4 Functional Checks 4 33 Section 4 6 Site Log Table 4 1 Vivid 4 Site Log Date Service Person Problem Comments...

Страница 184: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 4 34 Section 4 6 Site Log Table 4 1 Vivid 4 Site Log Continued Date Service Person Problem Comments...

Страница 185: ...his chapter contains block diagrams and functional explanations of the Vivid 4 electronic circuits Table 5 1 Contents in Chapter 5 Section Description Page Number 5 1 Overview 5 1 5 2 General Informat...

Страница 186: ...beam forming system which can handle up to 192 element linear probes by use of multiplexing Signal flow travels from the Probe Connector Panel to the Front End FE Electronics to the BEP and finally is...

Страница 187: ...ystem Block Diagram Vivid 4 BT03 Systems with RFI Configuration see Figure 5 1 below System Block Diagram Vivid 4 BT03 Systems with RFI Configuration see Figure 5 1 on page 5 3 System Cabling Diagram...

Страница 188: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 4 Section 5 1 Overview Figure 5 2 System Block Diagram Vivid 4 BT03 Systems with RFT Configuration PCVIC...

Страница 189: ...5 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2269429 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2269430 AC...

Страница 190: ...77 2252975 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2269430 A...

Страница 191: ...ents as described in the following sections Front Board Assembly FB on page 5 16 MUX Board on page 5 18 Beamformer Board BF on page 5 19 Radio Frequency Interface RFI Board on page 5 21 OR Front End C...

Страница 192: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 8 Section 5 1 Overview Figure 5 5 Front End Crate Block Diagram RFI Configuration...

Страница 193: ...I Radio Frequency Interface Board OR FEC Front End Controller Board RFT Radio Frequency Tissue Board IMP Image Port Front Board TR4 Boards MUX Board 64Ch Beam Former A D Board RFT Board Image Port Boa...

Страница 194: ...it bursts are generated on the Front board FB initiated by the transmit trigger pulse TXTRIG_L The transmit trigger starts the Transmit Pulse Generators TPG on the FB each generating 16 transmit pulse...

Страница 195: ...to 80 V HV1 another providing a voltage ranging from 0 to 40 V HV2 and the third outputting 80 V for the multiplexers in the linear probes HV1 and HV2 are programmable through a serial interface from...

Страница 196: ...EC or RFI and the FEC or RFI sends a signal to the high voltage power supply HVPS to program one of the supplies HVPS1 or HVPS2 Once the signal is transmitted it is received by the transmit pulsor TP...

Страница 197: ...ed delay between the elements and immediately after this are received all the time aiming at one point at a specific depth In this way echo signals are collected from all points along a certain line f...

Страница 198: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 14 Section 5 1 Overview Figure 5 8 Received Signal Path Block Diagram Systems with RFI Configuration...

Страница 199: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 5 Components and Function Theory 5 15 Figure 5 9 Received Signal Path Block Diagram Systems with RFT Configuration...

Страница 200: ...neral Description The FB Board has 64 identical receive channels The signal in each channel is fed via a Transmit Receive T R switch to a preamplifier The next step is a Time Controlled Gain TGC ampli...

Страница 201: ...nels Each channel has two pulsars for different transmitted Tx voltages and a low signal receiver amplifier Both are connected to the same input output line The receiver is buffered from the transmitt...

Страница 202: ...em to the BF Boards Expanding the low high channels in CW mode TEE motion control Monitoring temperature and voltage measurements via the I C FE Bus switch To reduce the noise in the FE cards the MUX...

Страница 203: ...tput from the other BA Level 2 as shown in Figure 5 13 on page 5 20 These ASICs will add all the input and output channels with appropriate delay to give optimal receive focusing and beam steering as...

Страница 204: ...1 Overview 5 4 4 2 Description of Operation Figure 5 13 Beamformer Board 64 Channel Block Diagram ADC ADC ADC ADC From prev board To next board BA2 FE bus 21 21 12 Bus Ifc RX N RX N 1 RX N 2 RX N 3 4...

Страница 205: ...y the RFI board System Master Clock generator Transmitter Control ATGC Control Receive Focusing Control Analog Test Signal Generator High Voltage Control and supervision Probe Sensing Control of MUX b...

Страница 206: ...clipping is performed 5 4 5 6 Color Flow Processing No Color Flow processing is performed here except for the RF demodulation previously mentioned The data used for Color Flow is tagged with the appr...

Страница 207: ...ollows TX Power Supply 1 TX1 generates voltage levels for pulsed TX operations TX Power Supply 2 TX2 generates voltage levels for both pulsed and CW operations The RFI sets the voltage level for the t...

Страница 208: ...A N A 9 2 5 V voltage Locally generated from 3 3 V 10 DSP CLOCK 50 MHz 11 15 V MONITOR voltage 12 EMIFA CLOCK 133 MHz 13 DSP_BSOE3_L signal 14 1 4 V voltage Locally generated from 3 3 V 15 OPTIONAL E...

Страница 209: ...voltage 35 5 V voltage 36 5 V voltage 37 N A 38 N A 39 2 0 V voltage 40 FAN VENT 1 voltage 41 FAN VENT 2 voltage 42 10V_Ref_voltage 43 5V_ref voltage 44 5V_ref voltage 45 AUDIO RIGHT CHANNEL OUT 1 V...

Страница 210: ...ter control ATCG control Receive focusing control Analog test signal control Probe sensing Transmit voltage control Control of Front Board switching Temperature sensing for the TEE probes Memory refre...

Страница 211: ...Bus 5 4 7 1 Control The RFT Board has a local DSP with an external EPROM The DSP performs the following control tasks Handles communication with the BE over the VME Bus and the PCI Bus Updates filter...

Страница 212: ...elow The IMP is used as a connection to the BEP All the instructions to the Front End Crate are sent via the IMP and all the outgoing data is transferred to the BEP via the IMP through the PCI Bus 5 4...

Страница 213: ...PC2IP on page 5 38 Plug and Scan Card and Battery on page 5 38 Network Card Onboard on page 5 38 SCSI Card on page 5 38 Floppy Drive on page 5 39 Hard Disk on page 5 39 Magneto Optical Drive MOD on p...

Страница 214: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 30 Section 5 1 Overview Figure 5 16 Back End Component Locations Vivid 4 BT03 Systems with RFI Configuration...

Страница 215: ...dio Bracket PCI2IP Card Multifunction I O Card Key Board Control Card MO Driver ECG Module CD R Driver Floppy Driver HD Driver Mother Board P S Battery Slot 1 Slot 2 Slot 3 Slot 4 Slot 5 Slot 6 AGP Sl...

Страница 216: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 5 32 Section 5 1 Overview Figure 5 18 Back End Block Diagram Vivid 4 BT03 Systems with RFI Configuration...

Страница 217: ...ug Scan Card SCSI Card Network Card PCI2IP Card Empty slot Key Board Control Card MO Driver ECG Module CD R Driver Floppy Driver HD Driver P S Battery Video Out Mother Board S VGA To Front End To LAN...

Страница 218: ...5 2 Central Processing Unit CPU The CPU controls and processes the internal Back End Processor operations In addition the CPU supports the Front End Crate via the IP card The CPU used in the Vivid 4 u...

Страница 219: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 5 Components and Function Theory 5 35 Figure 5 20 BEP Motherboard Layout Vivid 4 BT03 Note Not drawn to scale...

Страница 220: ...al Port Connector Header CPU Fan CPU Fan Header DIMM1 2 Memory DIMM Slots Ethernet LAN Ethernet Port Game Port Game Port J1a a P4SGA only 4xAGP J5 Floppy Disk Drive Connector J6 J7 IDE 1 2 Hard Disk D...

Страница 221: ...10 seconds This is for emergencies only and should not be used under regular circumstances Standby Mode which enables system boot up in 30 seconds this option is enabled if the Plug Scan option is ins...

Страница 222: ...PU The PCI2IP is sometimes referred to as the PCI Bridge 5 5 7 Plug and Scan Card and Battery The Plug Scan enables the Standby Mode during Plug and Scan operations and performs a legal shutdown of th...

Страница 223: ...the Vivid 4 ultrasound unit The Hard Disk is controlled by the CPU via signals from the IDE Bus The Hard Disk is partitioned into four drives each of which is designated for different operations as fo...

Страница 224: ...d to back up images and reports in the similar way as the MOD In addition the CD is used as the main source of software upgrades and other service utility operations The CDRW can write to 24X speed CD...

Страница 225: ...5 5 15 Modem The Modem is a universal modem controlled by the CPU via the USB port The modem is used for iLinq InSite remote servicing This is a standard MultiTech modem that connects to an analog pho...

Страница 226: ...image is captured and transmitted in RGB format to the S Video C Video Converter circuit and to the RGB to Color Printer Converter circuit The PC2IP board controls the PC VIC Power on and also contro...

Страница 227: ...Timing Interlaced Vertical Start VS 53 Lines field Active Lines VA Lines frame PAL standard Total Lines VT 625 Lines frame PAL standard Frame Rate 25Hz PAL standard Table 5 6 NTSC Video Specification...

Страница 228: ...ch receives and transmits Y C video signals is used for recording and re playing exams The appropriate type of VCR should be used for the local video signal PAL or NTSC 5 6 1 2 Black White Video Print...

Страница 229: ...components and converted into DC voltages in the BEP and the Front End to support their operation Table 5 7 below shows which voltages are used in the various system components both in the BEP and the...

Страница 230: ...e AC System see Figure 5 26 comprises the following components AC Isolation Transformer AC Distribution Box as described in AC Distribution Box on page 5 47 AC Input Box see AC Input Box on page 5 47...

Страница 231: ...yback video signal from the VCR to the Image Port 5 7 3 1 AC Input Box The AC Voltage Input cord is connected to the AC Input Box It contains the Dual Rating Circuit Breaker 6A 200 240V or 12A 100 120...

Страница 232: ...with specific requirements as indicated on the label near the AC input connector The AC transformer s primary and secondary voltages are set up by using the appropriate plugs connected in Jumper1 and...

Страница 233: ...RECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 5 Components and Function Theory 5 49 5 7 4 Front End DC Power Distribution Figure 5 28 DC Distribution Front End Crate Vivid 4 Systems wi...

Страница 234: ...PMX85N 80 80 40 40 5 10 PMX85P PMX85N DOLPHIN Front Bd MUX Bd BF2 BF1 FEC RFT Image Port J11 J21 J22 J14 J24 J25 J15 J16 J26 J33 J32 J31 J41 J42 J43 J53 J63 J73 J62 J52 J51 J61 J71 J72 125 125 125 125...

Страница 235: ...nt End itself Air is sucked from the inside of the system through the Filter Cover and the Filter It passes the Fan Assembly directly into the upper part of the Front End where it forms a cooling air...

Страница 236: ...th iLinq The following sections contain a brief introduction of iLinq s features 5 9 2 1 Web Server Browser The Service platform and other Service software use the iLinq web server and browser 5 9 2 2...

Страница 237: ...service software across all GEMS products 5 9 3 1 Internationalization The user interfaces provided by the service platform are designed for GE personnel and as such are in English only At this time...

Страница 238: ...Site use Every access request whether successful or not will be logged into a service access log that is viewable to authorized users See Chapter 7 Diagnostics Troubleshooting for more info about InSi...

Страница 239: ...iew 6 1 1 Purpose of Chapter 6 Table 6 1 Contents in Chapter 6 Section Description Page Number 6 1 Overview 6 1 6 2 Input AC Voltage Configuration 6 2 6 3 Front End Voltages and Signal Indicators 6 3...

Страница 240: ...ccording to specific requirements as indicated on the label near the AC input connector The AC transformer s primary and secondary voltages are set up by using the appropriate plugs connected in Jumpe...

Страница 241: ...e have red green and yellow LEDs which indicate the board s status as follows Red LED Lit An error has occurred on the board unless otherwise specified Yellow LED Lit Software defined Green LED Lit Th...

Страница 242: ...Front End Crate LEDs RFT Configuration The LEDs for each of the boards in the Front End Crate are summarized in the following subsections The LEDs are listed as viewed from the top of the board when...

Страница 243: ...Transmit data to BEP During Processing LD502 Red Not Used LD501 Yellow Not Used LD500 Green Not Used LD12 Green 5 0 V status LD22 Green 2 0 V status LD11 Green 1 4 V status LD10 Green 1 5 V status LD...

Страница 244: ...No Probe Lit if no probe is selected Yellow HV Over Range Lit if excessive HV is output possibly due to HV surveillance Table 6 6 RF and Tissue Processor LEDs LED Indicates When Lit Comments Red Board...

Страница 245: ...Normally lit Green 5V PS is OK Normally lit Table 6 9 Front Board Assembly LEDs LED Indicates When Lit Comments Green 15V LVPS is OK Normally lit Green 15V LVPS is OK Normally lit Green AVee LVPS is O...

Страница 246: ...Figure 6 3 Back End Power Supply Voltages Section 6 5 VIC Video Signal Setting The Vivid 4 ultrasound scanner and VIC may be configured to operate with either PAL or NTSC video systems as required 6 5...

Страница 247: ...may be accessed directly from the three adjustment buttons located on the front of the monitor as shown in Figure 6 4 below NOTE All display monitor controls other than the contrast and brightness con...

Страница 248: ...er to the recommended settings in Table 6 10 below 3 When the Contrast setting is correct press the center button again to make adjustments to the Brightness The Brightness adjustment bar is displayed...

Страница 249: ...ar scanning mode enables you to perform three types of calibration as described in the following sections Monitor Calibration section page 6 13 Beamformer Calibration section page 6 14 Video Grabbing...

Страница 250: ...DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 6 12 Section 6 7 Calibration 2 Trackball to the Calibration button and press Select The following calibration options are displayed Figure 6 6 C...

Страница 251: ...to the Display option and press Select The Calibration Display window is displayed as shown below Figure 6 7 Calibration Display Window 3 Adjust the brightness of the monitor so that the black square...

Страница 252: ...sk perform the BF calibration procedure 1 Access the calibration options as described in the Calibration section page 6 11 2 Trackball to the Calibration button and press Select The bias voltages for...

Страница 253: ...lmost the same quality as the left side Note The video signal path is from the Back End SVGA board output to the VIC via the rear panel connector IP and then back to the Back End via the PC2IP cable 5...

Страница 254: ...nd run diagnostic tools to locate system problems and failures The Vivid 4 ultrasound unit includes built in diagnostic tools that can be accessed from the scanner application Table 7 1 Contents in Ch...

Страница 255: ...emperatures in the Front End motherboard and in the AC Distribution Box To check the Front End parts the tests must follow a logical sequence in order to identify the faulty component For example test...

Страница 256: ...tools are accessed from 2D Mode by simultaneously pressing Ctrl F11 on the alphanumeric keyboard The Diagnostic menu is displayed as shown below Figure 7 1 Diagnostic Menu Note Once the test has been...

Страница 257: ...ing Front End FE Diagnostics section on page 7 5 Tests various back end hardware components as described in the Performing Back End Diagnostics on the System section on page 7 69 Monitors all low volt...

Страница 258: ...Diagnostic Test as described in the Beamformer BF Diagnostic Tests section on page 7 21 At this stage to continue to the Front End diagnostic tests it is necessary to re boot the system 3 Front Board...

Страница 259: ...manual options see Manual Options on page 7 8 7 3 1 1 Auto Sequence 1 In the Diagnostic menu trackball to the Auto Sequence button and press Select The Automatic FE Testing dialog box is displayed as...

Страница 260: ...under Execute Sequence and press Select if required 6 To resume testing from the point where you paused trackball to the Start button and press Select 7 At any stage during the testing sequence trackb...

Страница 261: ...Diagnostics 7 3 1 2 Manual Options 1 In the Diagnostic menu trackball to the Front End button and press Select The FE Diagnostics dialog box is displayed as shown in Figure 7 3 or Figure 7 4 below Fig...

Страница 262: ...DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 9 Figure 7 4 FE Diagnostics Dialog Box RFT Configuration Data Flow Map Test Description Area Test Area...

Страница 263: ...splays a description of the selected diagnostic test Test Area Displays the buttons that enable you to perform diagnostic tests Each button has an LED that indicates the following by its color Green T...

Страница 264: ...may be tested by performing either of the two tests Self Test or Full Test as required While the Full Test is in progress it may be necessary to wait a while since this requires longer than the Self T...

Страница 265: ...s contact your local GE Medical Systems field engineer for assistance If errors were identified but the results were marginal perform calibration as described in the Calibration section on page 6 11 a...

Страница 266: ...gnostics dialog box as described in the Accessing the Front End Diagnostic Options section on page 7 6 2 In the FE Diagnostics dialog box trackball to the IMP tab and press Select OR Trackball to the...

Страница 267: ...Comment area and proceed as follows If no problems were identified proceed to the VME test as described in the VME Bus VME Diagnostic Tests for RFT Configuration section on page 7 15 If no problems we...

Страница 268: ...section on page 7 13 If the VME Bus test is performed without first performing this test and errors are identified you must perform the IMP test to ensure that the VME board is the source of the error...

Страница 269: ...area and proceed as follows If no problems were identified proceed to the RFT test as described in the RFT Diagnostic Tests for RFT Configuration section on page 7 17 If no problems were identified bu...

Страница 270: ...ibed in the Image Port IMP Diagnostic Tests for RFT Configuration section on page 7 13 and in the VME Bus VME Diagnostic Tests for RFT Configuration section on page 7 15 If the RFT test is performed w...

Страница 271: ...ere identified during the performance of the test 5 Review the test results in the Comment area and proceed as follows If no problems were identified proceed to the FEC test as described in the Front...

Страница 272: ...Temp Test Diagnoses the temperature threshold for TEE probes Pencil ID Test Checks the pencil ID circuit in the FEC Note Before performing any of the FEC tests perform the IMP VME Bus and RFT tests as...

Страница 273: ...dently The relevant LED changes color as follows Yellow The test is in progress Green The test has been completed and no problems were identified Red Problems were identified during the performance of...

Страница 274: ...the VME Bus test and the IMP test to ensure that the BF board is the source of the error NOTE When replacing the BF perform calibration as described in the Calibration section on page 6 11 NOTE IMPOR...

Страница 275: ...N 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 22 Section 7 3 Performing Front End FE Diagnostics Figure 7 11 BF Tab FE Diagnostics Dialog Box RFI Configuration Data Flow Map Test Description Area...

Страница 276: ...ital Test 1 independently as follows Trackball to the Digital Test 0 button and press Select When complete trackball to the Digital Test 1 button and press Select 4 In the BF tab verify that the Comme...

Страница 277: ...lternatively review the information manually using the scroll arrows on the right of the Ultrasound Scope window to select a vector between 0 and 63 The selected vector is displayed in the Vector fiel...

Страница 278: ...neer for assistance If errors were identified but the results were marginal perform calibration as described in the Calibration section on page 6 11 and then repeat the test If errors were identified...

Страница 279: ...11 If the FB tests are performed without first performing these tests and errors are identified you must perform the following tests in the order listed to ensure that the FB is in fact the source of...

Страница 280: ...CTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 27 Figure 7 14 F B Tab FE Diagnostics Dialog Box RFI Configuration Data Flow Map Test Description Area Test...

Страница 281: ...Test Checks the transmit switches as described in the TR Switch Test section on page 7 39 d FLA Jumper Test Checks the jumpers that connect the side elements to the center elements as described in th...

Страница 282: ...test is in good condition 1 Connect either a 7S or 3S probe and lock it in place 2 Access the F B tab of the FE Diagnostics dialog box as described in the Accessing the Front Board Assembly Options se...

Страница 283: ...he test result message in the Report dialog box If one or more channels were identified as bad trackball to the Save Report button and press Select then contact a GE Medical Systems service representa...

Страница 284: ...esults it is imperative that only this probe is used and that it be in good condition 2 Access the F B tab of the FE Diagnostics dialog box as described in the Accessing the Front Board Assembly Optio...

Страница 285: ...eckbox to display only bad channels in the Display area Mark bad channel Select this checkbox to manually mark the currently selected channel as bad 6 Check the test result message and proceed as foll...

Страница 286: ...ard elements as follows Each element is represented by a button highlighted in green that displays the element and the channel number that is associated with that element Each button is highlighted ye...

Страница 287: ...s Select and contact a GE Medical Systems service representative If no errors were identified proceed with the next test If no errors were identified but the problem persists contact your local GE Med...

Страница 288: ...t is required if the tests are not being performed in the sequence described in the Performing Front End FE Diagnostics section on page 7 5 1 Disconnect all the probes 2 Access the F B tab of the FE D...

Страница 289: ...Show Only Bad Channels Select this checkbox to display only bad channels in the Display area Mark bad channel Select this checkbox to manually mark the currently selected channel as bad 5 Check the t...

Страница 290: ...p consists of four TR4 sockets each socket containing four vectors channels as follows Each vector is represented by a button highlighted in green that displays the number of the socket to which the T...

Страница 291: ...s Select and contact a GE Medical Systems service representative If no errors were identified proceed with the next test If no errors were identified but the problem persists contact your local GE Med...

Страница 292: ...B tab of the FE Diagnostics dialog box as described in the Accessing the Front Board Assembly Options section on page 7 26 3 In the Select Socket area of the F B tab trackball to the socket to which t...

Страница 293: ...els were identified as bad trackball to the Save Report button and press Select then contact a GE Medical Systems Service Representative If no channels were identified as bad mark the channels suspect...

Страница 294: ...has two phases namely the open pencil probe test and the close pencil probe test 1 Connect a 2D pencil probe 2 Access the F B tab of the FE Diagnostics dialog box as described in the Accessing the Fr...

Страница 295: ...channels in the Display area Mark bad Channel Select this checkbox to manually mark the currently selected channel as bad 5 Check the test result message and proceed as follows If one or more relays...

Страница 296: ...of the eight front board pencil probe relays as follows Each relay is represented by a button highlighted in green that displays the relay number Each button is highlighted in yellow when selected Eac...

Страница 297: ...rt and press Select and contact a GE Medical Systems service representative If no errors were identified proceed with the next test If no errors were identified but the problem persists contact your l...

Страница 298: ...are displayed in the Comments area 4 Check the test results and proceed as follows If no errors were identified proceed with the next test If no errors were identified but the problem persists contac...

Страница 299: ...m focal beam Note You must reboot the system after performing the RAM test 1 Access the F B tab of the FE Diagnostics dialog box as described in the Accessing the Front Board Assembly Options section...

Страница 300: ...ists contact your local GE Medical Systems field engineer for assistance If errors were identified but the results were marginal perform calibration as described in the Calibration section on page 6 1...

Страница 301: ...Options 1 Access the FE Diagnostics dialog box as described in the Accessing the Front End Diagnostic Options section on page 7 6 2 In the FE Diagnostics dialog box trackball to the MUX tab and press...

Страница 302: ...CTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 49 Figure 7 27 MUX Tab FE Diagnostics Dialog Box RFT Configuration Data Flow Map Test Description Area Test...

Страница 303: ...l on page 7 59 AC Power Input Test on page 7 61 LVPS Voltage Test on page 7 63 HVPS Voltage Test on page 7 64 Temp Test on page 7 65 Show History on page 7 66 7 3 9 2 Normal Channels Diagnostic Tests...

Страница 304: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 51 Figure 7 28 Normal Channels Diagnostic Test...

Страница 305: ...Bad Channels Select this checkbox to display only bad channels in the Display area Mark bad channel Select this checkbox to manually mark the currently selected channel as bad 4 Check the test result...

Страница 306: ...group consists of four TR4 sockets each socket containing four vectors channels as follows Each vector is represented by a button highlighted in green that displays the number of the socket to which t...

Страница 307: ...s Select and contact a GE Medical Systems service representative If no errors were identified proceed with the next test If no errors were identified but the problem persists contact your local GE Med...

Страница 308: ...ting 7 55 7 3 9 3 Expanded Low Channels Diagnostic Tests Tests output signals in CW Mode only when the input channels are connected to the output channels If one channel is missing the output will be...

Страница 309: ...orming Front End FE Diagnostics 7 3 9 4 Expanded High Channels Diagnostic Tests Tests the output signals from the input of the high channels If one channel is missing the output will be 6db The input...

Страница 310: ...SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 57 7 3 9 5 Port Test The Port test is a general logic control test Figure 7 32 Port Test RFI Con...

Страница 311: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 58 Section 7 3 Performing Front End FE Diagnostics Figure 7 33 Port Test RFT Configuration...

Страница 312: ...ll as temperatures 7 3 9 6 1 TEE Probe Control The TEE probe control test checks the position setting of the probe as well as its normal operation 1 Access the MUX tab of the FB Diagnostics dialog box...

Страница 313: ...ECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 60 Section 7 3 Performing Front End FE Diagnostics Figure 7 35 TEE Probe Test Dialog Box 5 Verify the test results and then trackball to the OK b...

Страница 314: ...accuracy by percentage 1 Access the MUX tab of the FB Diagnostics dialog box as described in the Accessing the MUX Diagnostic Test Options section on page 7 48 2 Trackball to the Advanced board tests...

Страница 315: ...nt End FE Diagnostics 4 Trackball to the AC Power Input button and press Select The AC Power Input dialog box is displayed Figure 7 37 AC Power Input Dialog Box 5 Trackball to the Show button and pres...

Страница 316: ...X Diagnostic Test Options section on page 7 48 2 Trackball to the Advanced board tests button and press Select The Advanced System Tests options are displayed as shown in Figure 7 34 on page 7 59 3 Tr...

Страница 317: ...ox as described in the Accessing the MUX Diagnostic Test Options section on page 7 48 2 Trackball to the Advanced board tests button and press Select The Advanced System Tests options are displayed as...

Страница 318: ...section on page 7 48 2 Trackball to the Advanced board tests button and press Select The Advanced System Tests options are displayed as shown in Figure 7 34 on page 7 59 3 Trackball to the Monitor but...

Страница 319: ...Advanced board tests button and press Select The Advanced System Tests options are displayed as shown in Figure 7 34 on page 7 59 3 Trackball to the Monitor button and press Select The View Power opti...

Страница 320: ...the Front End boards software versions and hardware revisions 1 Access the FE Diagnostics dialog box as described in the Accessing the Front End Diagnostic Options section on page 7 6 2 Trackball to...

Страница 321: ...n the FE Diagnostics dialog box enables you to generate a report of the current test and failure results 1 Access the FE Diagnostics dialog box as described in the Accessing the Front End Diagnostic O...

Страница 322: ...End hardware components Note The Back End Diagnostic Tests may be performed independently regardless of sequence 7 4 1 Accessing the Back End Diagnostic Options 1 In 2D Mode press Ctrl F11 on the alph...

Страница 323: ...tions as described in the ECG Phono Diagnostic Test section on page 7 72 External Keyboard Enables you to test the keyboard options and lights in different modes as described in the External Keyboard...

Страница 324: ...rmed for each of the speakers at various PRFs and frequencies Left and right are as seen when standing behind the unit 3 In the User controls area trackball to the Right radio button and press Select...

Страница 325: ...r to an appropriate ECG simulator In this case select the External ECG radio button in the TestECGDevice dialog box 2 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnost...

Страница 326: ...n page 7 69 3 Trackball to the ECG Phono button and press Select The TestECGDevice dialog box is displayed as shown in Figure 7 47 on page 7 72 4 In the Select trace area trackball to the Phono radio...

Страница 327: ...iagnostic Test section on page 7 77 7 4 4 1 Manual External Keyboard Test 1 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnostic Options section on page 7 69 2 Trackbal...

Страница 328: ...ector on the front panel of the unit and press the right middle and left of the switch Verify that the numbers 68 69 and 70 accordingly are displayed on the info bar 11 When you are finished trackball...

Страница 329: ...al radio button and press Select 4 In the KB Panel tab deselect the Select checkbox Verify that after the Select LED is off the alphanumeric keyboards LEDs are dimmed 5 When you are finished trackball...

Страница 330: ...nality of the alphanumeric keyboard keys 1 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnostic Options section on page 7 69 2 Trackball to the Keyboard button and pres...

Страница 331: ...splayed as shown below Figure 7 52 Util_MO_CDR_App Dialog Box 3 Trackball to the MO radio button and press Select 4 Trackball to the IsMediaAccess button an press Select The message Media in drive G i...

Страница 332: ...d 7 Select No The message Media is outside is displayed Verify that the media has actually been ejected 8 Trackball to the Load button and press Select The message Media is inside is displayed Verify...

Страница 333: ...s check the memory usage 7 4 7 1 Benchmark Testing 1 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnostic Options section on page 7 69 2 Trackball to the Computer Testi...

Страница 334: ...h indicates current computer results are higher or equal to those indicated in the lower line 7 4 7 2 Check Disk Bus 1 Access the HW Diagnostics menu as described in the Accessing the Back End Diagnos...

Страница 335: ...Bench In the information displayed make sure that the speed values displayed are greater than 23MB sec This test checks the speed of information transfer between the hard disk and IDE BUS Bus Bench In...

Страница 336: ...t The system enters Standby mode The light on the ON OFF button on the control console blinks repeatedly 3 Disconnect the AC power cable from the mains wall power outlet 4 Check that the ON OFF button...

Страница 337: ...allel or USB port on the Vivid 4 rear connectors panel 1 Turn ON the system by pressing the ON OFF button on the control console and re boot into Technical Support Mode using the T plug While re booti...

Страница 338: ...Connectivity Provides basic connectivity between the scanner and the OnLine Center OLC or the field engineer in the field Configuration Provides the interface to configure the various iLinq parameter...

Страница 339: ...al Support tab as shown in Figure 7 58 below Figure 7 58 Technical Support System Configuration 2 Trackball to OK and press Select The System Configuration window closes Note If iLinq was not found to...

Страница 340: ...location select Remote from the Profile Name drop down list 5 Trackball to the Start Communicator button and press Select Wait a while until the Service Login dialog box is displayed as shown below Fi...

Страница 341: ...ct the required tool or application as described in the following sections Error Logs Page section on page 7 90 Diagnostics Page section on page 7 98 Image Quality Page section on page 7 111 Calibrati...

Страница 342: ...Logs Page When the Error Logs page is selected as shown in Figure 7 62 below the log viewing options described in the following sections are available Log Viewer section on page 7 91 Log Browser sect...

Страница 343: ...e following high level requirements Simple filtering of the scanner log s with filtering capabilities as a function of login access permissions Log visibility by all services modes Multiple instances...

Страница 344: ...he power logs including errors monitoring dates and values as shown below Figure 7 65 Power Logs Temperature Displays all the temperature logs including errors monitoring dates and values as shown bel...

Страница 345: ...Power logs in graphical format Select Power or Temperature from the Log option on the Logs Viewer menu and then select Plot Log to display the results as shown below Figure 7 67 Plot Log Window Plot...

Страница 346: ...sitive text that you want to filter from the logs or pages that you are currently viewing as shown below Figure 7 69 Search for Text in Log or Page 7 5 3 1 4 Filter The Filter option on the Logs Viewe...

Страница 347: ...Viewer Select the Exit Log Viewer option to return to the Common Service Desktop Figure 7 71 Exit Option 7 5 3 2 Log Browser When Log Browser is selected in the left pane of the Error Logs page you ca...

Страница 348: ...mmon Service Interface Each system re boot will generate a new log file each log file can be viewed by selecting the appropriate link as shown in the example in Figure 7 73 below Figure 7 73 Viewing a...

Страница 349: ...m this screen Figure 7 73 i e when you are connected to the system remotely or you may save the file to a floppy disk on the system itself Only three log files may be reviewed in the field DiagErr Sys...

Страница 350: ...teractive The user is required to perform an operation on the ultrasound unit in order for the test to be completed successfully This option is not applicable when used remotely Non Interactive The te...

Страница 351: ...rey abort Execute Button Enables you to start a selected diagnostic test Once the diagnostic is running the Execute button becomes the Abort button enabling you to stop the diagnostic Loop Count Field...

Страница 352: ...AL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 100 Section 7 5 Common Service Interface Reset Power Supply Enables you to reset the Front End power supply Figure 7 76 Reset Power...

Страница 353: ...ON 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 101 Restart Scanner Software System Shutdown Enables you to remotely shutdown or restart the system Figure 7 77...

Страница 354: ...51 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 102 Section 7 5 Common Service Interface Save System State to Log This performs the same system operation as that initiated when pressing Ctrl F or Alt D Fig...

Страница 355: ...d CPU String Operations Tests the proper operation of the string commands LODS MOVS SCAS and STOS with 8 16 and 32 bit operands If errors are detected the failed commands and operands are listed CPU L...

Страница 356: ...ing these diagnostics will also look for parity errors and other exceptions If encountered these are added to the test log as errors Memory Pattern Test Uses several test patterns to test as much memo...

Страница 357: ...nd video controller 7 5 4 3 6 Frame Grabber Test Testing for proper operation of the Frame Grabber Board is achieved by performing the Video Grabbing Test For details refer to the Video Grabbing Test...

Страница 358: ...0 Monitor Test Combination Test Checks the monitor alignment color depth and monitor resolution Four cross hairs are placed in each corner of the screen as well as RGB spectrum arrays and horizontal a...

Страница 359: ...s as those performed by pressing Ctrl F11 during regular operation can be conducted Both Non interactive see Figure 7 81 or Figure 7 82 and Interactive see Figure 7 83 tests may be performed However t...

Страница 360: ...for RFI Configuration on page 7 11 OR for systems with RFT Configuration Image Port Image Port IMP Diagnostic Tests for RFT Configuration on page 7 13 VME Bus VME Bus VME Diagnostic Tests for RFT Con...

Страница 361: ...DICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 109 Figure 7 83 Interactive Tests NOTE Interactive tests are performed on the Front Board...

Страница 362: ...est refer to the appropriate section as follows Transmit Test see Transmit Test on page 7 29 FLA Jumper Test see FLA Jumper Switch on page 7 31 TR Switch Test see TR Switch Test on page 7 39 Note On t...

Страница 363: ...as shown below you can verify and calibrate image quality Figure 7 84 Image Quality Page Note This page is not populated in this version 7 5 6 Calibration Page When the Calibration page is selected as...

Страница 364: ...ge When the Configuration page is selected you can configure the GSUI InSite No changes should be made in this page without the help of GE Medical Systems Service Personnel as any modifications made b...

Страница 365: ...size already utilized IP Configuration shows the IP configuration of the different remote devices such as the network or the modem Network Status shows the communication status of active connections...

Страница 366: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 7 114 Section 7 5 Common Service Interface Figure 7 88 System Shutdown...

Страница 367: ...tem Shutdown There are three options provided described on previous page also see Figure 7 88 Reset Database used to empty the database and reset a new one see Figure 7 89 below Figure 7 89 Reset Data...

Страница 368: ...gs CAUTION THIS PROCESS IS IRREVERSIBLE Before proceeding make sure you have a complete backup of the current settings if they will be required in the future When selecting Clean Userdefs the followin...

Страница 369: ...Select the Prodiags Browser under the Configuration Utility option The following message is displayed Wait a few minutes until the following screen is displayed Figure 7 91 ProDiags Browser Schedule...

Страница 370: ...EVISION 3 VIVID 4 SERVICE MANUAL 7 118 Section 7 5 Common Service Interface Figure 7 92 ProDiags Browser Task The Task tab indicates all the tasks that are available for activation This also indicates...

Страница 371: ...TEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 119 Figure 7 93 ProDiags Browser Time Slot The Time Slot tab indicates the time slots available for...

Страница 372: ...L 7 120 Section 7 5 Common Service Interface Figure 7 94 ProDiags Browser Log and Result The Log Result tab indicates the logs and results of any tasks that have been performed At any of the above men...

Страница 373: ...NUAL Chapter 7 Diagnostics Troubleshooting 7 121 7 5 8 3 2 File Transfer Configuration The File Transfer Configuration utility can be used to transfer files to the GE Service center Follow the on scre...

Страница 374: ...terface 7 5 8 3 3 iLinq Configuration Utility When the iLinq Configuration Utility option is selected in the left pane of the Configuration page see Figure 7 96 below the InSite Interactive Platform C...

Страница 375: ...3 VIVID 4 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 123 Click Accept and select the modem device as the communication device to the OLC in the Device Connection tab also shown below Figu...

Страница 376: ...p information required for the InSite Checkout tab as shown below When all the settings have been entered click the CHECKOUT NOW button twice The OLC will attempt to communicate with your ultrasound u...

Страница 377: ...under the Technical Support Tab will allow you to save the iLinq IP settings to the hard disk as described below iLinq IP Save Procedure 1 In the System Configuration window click on the Technical Su...

Страница 378: ...n window under the Technical Support Tab refer to Figure 7 99 above trackball to iLinq Settings Restore and press Select A blank screen appears wait for a while until the phone and modem options messa...

Страница 379: ...tails and ordering information is displayed as described in Chapter 9 Renewal Parts Figure 7 100 Replacement Page Note This page is not populated in this version 7 5 10 PM Page When the PM page is sel...

Страница 380: ...ity logging you can also enter your own comments as well as screen captures pertaining to system functionality These comments are saved as part of the daily logger report 1 When an error occurs press...

Страница 381: ...ou expand the dropdown list in the Store Daily History area as shown below By default the current date is displayed Figure 7 103 Failure Report Calendar 3 Insert an empty PC formatted diskette Note An...

Страница 382: ...252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2269429 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2269430 AC DI...

Страница 383: ...53016 2252977 2252975 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On of...

Страница 384: ...his chapter provides replacement procedures for system parts Table 8 1 Contents in Chapter 8 Section Description Page Number 8 1 Overview 8 1 8 2 Cover Replacement Procedures 8 2 8 3 Control Console C...

Страница 385: ...2 Vivid 4 Ultrasound Unit Covers Legend to Figure 8 1 Label Item Label Item 1 Right Side Cover 7 Control Console Lower Cover 2 Left Side Cover 8 Control Console Upper Cover Front 3 Front Cover 9 Cont...

Страница 386: ...power unit turn the power OFF and disconnect the power cord CAUTION Before removing circuit boards turn the power OFF and wear the ESD wrist band strap 8 2 1 2 Order of Cover Removal Remove the side c...

Страница 387: ...rs four in total an example of which is shown below To easily access the screws turn the wheels inwards by 90 degrees Figure 8 2 Wheel Arch Screw 2 Slide the right side cover down and outwards until i...

Страница 388: ...e Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 2 3 3 Front Cover Removal Procedure 1 Pull out and remove the air filter as shown below Figure 8 3 Air Filter 2 Remove the side cover...

Страница 389: ...nit taking care to first insert its upper edge under the top cover 2 Fasten the front cover with the three screws at the edge of the top cover as shown in Figure 8 5 above 3 Fasten the left and right...

Страница 390: ...that secure the rear cover in position Two screws are located on the sides of the rear cover as shown below The other two screws are located in the AC Input Box Unit as shown in Figure 8 7 below Figu...

Страница 391: ...edures 8 2 5 2 Preparations Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 2 5 3 Connector Panels Cover Removal Procedure 1 Remove the two screws located one on either...

Страница 392: ...ed in the Side Covers Replacement Procedures on page 8 4 2 Remove the rear cover as described in the Rear Cover Replacement Procedures on page 8 7 3 Raise the control console to its maximum height 4 R...

Страница 393: ...lug any cables from below the rear panels except for those indicated in this procedure 8 Viewing the unit from behind there are two small metal plates at the back of the top cover each with two screws...

Страница 394: ...f the gas spring cover shown below and lower it into the collar of the top cover Figure 8 14 Gas Spring Cover Screws CAUTION The remaining section of the gas spring cover cannot be removed at this sta...

Страница 395: ...sten it with the two screws shown in Figure 8 14 on page 8 11 7 Gently fit the left and right connector panels back to their original position on the top cover and fasten each with its four screws 8 M...

Страница 396: ...s described in Chapter 3 Installation 8 2 7 3 Gas Spring Cover Removal Procedure The procedure for removing the gas spring cover is identical to the procedure for removing the lower section of the top...

Страница 397: ...console lower rear cover as described in the Control Console Lower Rear Cover Removal Procedure on page 8 19 2 Loosen the screws fastening the control console lower cover in position and lower the fr...

Страница 398: ...rear of the upper front cover refer to Figure 8 15 to fasten the front cover to the upper rear cover and tighten the 2 screws 4 Return and fasten the 12 screws 6 on each side and 2 in the front as sho...

Страница 399: ...bed in the Vivid 4 17 Monitor Removal on page 8 24 CAUTION The Vivid 4 17 monitor is heavy At least two people must be available to lift the monitor Attempts to move this unit by one person alone coul...

Страница 400: ...urn the control console lower cover as described in the Control Console Lower Cover Installation Procedure on page 8 18 5 Remove the control console lower rear cover as described in the Control Consol...

Страница 401: ...the Control Console Lower Rear Cover Removal Procedure on page 8 19 4 Remove the 16 screws 7 from each side and 2 from the front that fasten the lower cover in place as shown in Figure 8 17 below Figu...

Страница 402: ...unit as described in Chapter 3 Installation 8 2 11 3 Control Console Lower Rear Cover Removal Procedure 1 Raise the control console to its maximum height 2 Remove the 4 screws that fasten the rear co...

Страница 403: ...cribed in the Control Console Upper Cover Front Removal Procedure on page 8 14 3 Remove the two 8mm securing nuts two on each side and one nut located below the keyboard parallel to the hand rest loca...

Страница 404: ...as described in the Control Console Upper Cover Rear Removal Procedure on page 8 16 3 Remove the eight Allen 8mm securing screws four on each side from the rear of the upper console 4 Remove the Rear...

Страница 405: ...aker Replacement Procedure on page 8 34 8 3 1 Vivid 4 17 Monitor Rear Cover Replacement Procedure NOTE The procedures below describe replacement of the Vivid 4 Monitor Rear Cover only P N 2303931 8 3...

Страница 406: ...e the Vivid 4 control console is lowered to its minimum height 2 Carefully lift the replacement Monitor Rear Cover into position on the monitor 3 Return and fasten each of the 4 screws previously remo...

Страница 407: ...ures 8 3 2 2 Preparation Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 3 2 3 Vivid 4 17 Monitor Removal 1 Turn OFF the Vivid 4 main circuit breaker located at the rear...

Страница 408: ...n place before you return the monitor to the control console 1 Using two people lift the monitor into position on the control console 2 Fasten the monitor securing bracket in position using the four s...

Страница 409: ...rocedure on page 8 19 2 Remove the control console lower cover as described in the Control Console Lower Cover Removal Procedure on page 8 18 3 Remove the control console upper cover front as describe...

Страница 410: ...r refer to Figure 8 25 on page 8 26 3 Return and fasten the two M4 screws to secure the keyboard matrix to the metal frame one on each of the left and right corners refer to Figure 8 84 on page 8 99 4...

Страница 411: ...its neighboring keypad see Figure 8 26 below 2 Use the screw driver to gently lever off the defective keypad until it is completely loosened from its base 3 Lift off and remove the defective keypad NO...

Страница 412: ...allation 8 3 5 3 Keycaps Removal Procedure 1 Carefully place a small flat screw driver between the ATGC slider keycap you wish to remove and the surface of the keyboard 2 Use the screw driver to gentl...

Страница 413: ...nob replacement procedure 8 3 6 2 Preparation Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 3 6 3 Rotary Knob Removal Procedure 1 Carefully place a small flat screw dr...

Страница 414: ...the curved surface facing towards the center of the hole as shown in Figure 8 29 below Figure 8 29 Tension Plate Fitted in Centre of Knob 2 Carefully place the new rotary knob into position on the se...

Страница 415: ...n the Control Console Upper Cover Front Removal Procedure on page 8 14 4 Unplug the four keyboard cables located at the bottom of the keyboard see Figure 8 24 on page 8 26 5 Loosen and remove the two...

Страница 416: ...e left and right corners see Figure 8 25 on page 8 26 5 Reconnect the four keyboard cables to the keyboard previously disconnected see Figure 8 24 on page 8 26 6 Return the control console upper cover...

Страница 417: ...Console Upper Cover Front Removal Procedure on page 8 14 3 Remove the four speaker screws one from each corner of both of the speakers 4 Gently pull the speakers towards you 5 Disconnect the wires at...

Страница 418: ...ECTION POINT LOCATED ON THE REAR OF THE SCANNER TO THE RIGHT OF THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT 8 4 1 2 Preparation Shut down the Vivi...

Страница 419: ...frame guiding tracks 2 Use the plastic latches to firmly lock in the card in position so that the connectors are firmly placed in their sockets 3 Return the front door of the crate to its original po...

Страница 420: ...CATED ON THE REAR OF THE SCANNER TO THE RIGHT OF THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT 8 4 2 2 Preparation Shut down the Vivid 4 ultrasound...

Страница 421: ...et previously removed that secures the TR4 boards to its original location above the TR4 boards 3 Return and tighten the four screws two on each side to fasten the bracket in position and secure the T...

Страница 422: ...p and disconnect the AC power supply cable from the bottom of the DC power supply 3 Disconnect the AC power supply cable that is connected to the Tx P S HV P S 4 Remove the four small screws located o...

Страница 423: ...ove the DC power supply refer to the DC Power Supply Removal Procedure on page 8 39 3 Remove the two screws located at the top of the Tx power supply unit 4 Remove the Tx power supply unit by pulling...

Страница 424: ...probe sockets 3 Remove the right side cover as described in the Side Covers Removal Procedure on page 8 4 4 Remove the front door of the crate by unscrewing the two large screws at the top of the door...

Страница 425: ...he front end crate back to its original position 5 Fasten the front end crate to the frame with the eight screws 6 Reconnect the PCI cable to the Image Port Board or RFI Board and slide it back to its...

Страница 426: ...l Procedure 1 Remove the front end crate as described in the Front End Crate Removal Procedure on page 8 41 2 Unscrew and remove the four screws that secure the fans to the top of the front end crate...

Страница 427: ...gure 8 38 on page 8 43 CAUTION Ensure that you position the fan in the correct airflow direction pushing air inwards into the crate and that you connect the fan to the connector that was marked during...

Страница 428: ...paration Furnished Materials Confirm that the Back End Processor Kit P N 2315842 7 contains the following items Important The BEP arrives inside packing box P N 2328455 Make sure the returned BEP is s...

Страница 429: ...only the following components ready installed CD R driver Floppy driver HD driver Cables and Brackets Memory 512 MB Therefore the FE will need to remove some components from the removed BEP and instal...

Страница 430: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 8 Replacement Procedures 8 47 Figure 8 40 Back End Component Locations Systems with RFT Configuration...

Страница 431: ...rchive 2 Press Utility and select Backup 3 Start full incremental backup onto the removable media 4 When the backup procedure is completed the yellow portion of the pie chart will be 0 8 5 2 3 Back En...

Страница 432: ...KING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER TO THE RIGHT OF THE POWER CONNECTOR 2 FOLLOW...

Страница 433: ...in the MO Drive Installation Procedure on page 8 69 if applicable 2 Remove the Octopus Card Holder as described in the BEP2 Octopus Card Holder Removal Procedure on page 8 56 3 Install the SCSI Board...

Страница 434: ...on page 8 51 13 Connect the VIC VGA OUT Left cable Part No 2252964 to the VGA OUT socket 1 on the BEP rear panel 14 Connect the VIC VGA IN Right cable Part No 2252965 to the VIC VGA IN socket 2 15 Co...

Страница 435: ...cable Part No 2307289 to the Plug and Scan socket 16b and then connect the BEP Modem power cable Part No 2300847 to the power input socket 16a 31 Facing the side cover push the BEP as far to the right...

Страница 436: ...Fan Set to ON checkbox activated UPS Change the UPS setting to Used or Not Used as appropriate If the Plug Scan option was purchased the UPS setting can be selected as Used If this option not purchas...

Страница 437: ...cedure 1 Remove the two screws that secure the side cover in position from the rear of the BEP see Figure 8 44 below 2 Slide the cover backwards and remove it 8 5 3 4 BEP2 Side Cover Installation Proc...

Страница 438: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 8 Replacement Procedures 8 55 Figure 8 45 Front Cover BEP2 Figure 8 46 Front Cover Removal BEP2...

Страница 439: ...cover in position see Figure 8 45 on page 8 55 8 5 3 7 BEP2 Octopus Card Holder Removal Procedure 1 Remove the three screws that secure the Octopus Card Holder in position as shown in Figure 8 47 belo...

Страница 440: ...al Procedure on page 8 56 5 Disconnect all cables connected to the Plug Scan board from the keyboard control card the battery and the power supply Note Counting from the top the board is positioned in...

Страница 441: ...d unit as described in Chapter 3 Installation Voltage may exist in the system when a Plug Scan is present Ensure that a full shutdown of the system is performed otherwise voltage might be present that...

Страница 442: ...the BEP Figure 8 50 Removing Plug Scan Velcro Strips 1 Place the battery inside the battery bracket Ensure that the cable is positioned at the lower end and that the rear of the battery with the two v...

Страница 443: ...BEP 3 Remove the BEP side cover as described in the BEP2 Side Cover Removal Procedure on page 8 54 4 Remove the octopus card holder as described in the BEP2 Octopus Card Holder Removal Procedure on pa...

Страница 444: ...Procedure on page 8 4 2 Disconnect the power cable from the rear of the BEP 3 Remove the BEP side covers as described in the BEP2 Side Cover Removal Procedure on page 8 54 4 If applicable remove the o...

Страница 445: ...5 Back End Parts Replacement 7 Return the Vivid 4 rear and left and right side covers see the Rear Cover Installation Procedure on page 8 7 and the Side Covers Installation Procedure on page 8 4 Note...

Страница 446: ...5 Disconnect the PCI cable that is connected to the BEP Note Counting from the top the PC2IP board is positioned in the fourth slot 6 Release the securing screw located on the left of the board bracke...

Страница 447: ...val Procedure on page 8 56 5 Disconnect the frame grabber cable from the rear of the frame grabber board Note Counting from the top the frame grabber board is positioned in the fifth PCI slot 6 Releas...

Страница 448: ...ables connected to the keyboard control board Note Counting from the top the keyboard control board is positioned in the sixth PCI slot 6 Release the securing screw located on the left of the board br...

Страница 449: ...nstallation 8 5 11 3 CDRW Drive Removal Procedure 1 Remove the Vivid 4 left and right side covers and rear cover see the Rear Cover Removal Procedure on page 8 7 and the Side Covers Installation Proce...

Страница 450: ...from the drive previously removed 2 Carefully slide the CDRW Drive into the front of the BEP refer to Figure 8 51 on page 8 66 3 Reconnect the cables data and power to the rear of the CDRW Drive 4 Ins...

Страница 451: ...2 Remove the Vivid 4 left and right side covers refer to the Side Covers Removal Procedure on page 8 4 3 Remove the Vivid 4 rear cover refer to the Rear Cover Removal Procedure on page 8 7 4 Remove t...

Страница 452: ...CSI card see SCSI Board Installation Procedure on page 8 61 4 Connect the data cable between the SCSI card and the rear of the MO Drive 5 Connect the power cable to the rear of the MO Drive Make sure...

Страница 453: ...id 4 left and right side covers refer to the Side Covers Removal Procedure on page 8 4 3 Remove the Vivid 4 rear cover refer to the Rear Cover Removal Procedure on page 8 7 4 Disconnect all cables fro...

Страница 454: ...le in position two from each side of the exposed BEP as shown in Figure 8 54 on page 8 71 9 Disconnect the footswitch cable from the KB Controller card 10 Disconnect the COM2 cable from the rear of th...

Страница 455: ...COM 2 cable to the rear of the new ECG Module 5 Install the BEP left and right side covers as described in the BEP2 Side Cover Installation Procedure on page 8 54 6 Carefully return the BEP to its ori...

Страница 456: ...und unit as described in Chapter 3 Installation 8 5 14 3 PC VIC Removal Procedure 1 Remove the left cover as described Side Covers Removal Procedure section page 8 4 2 Remove the BEP side cover as des...

Страница 457: ...gure 8 57 above 3 Reconnect all the cables at the rear of the PC VIC module and the hard disk 4 Carefully slide the PC VIC into position in the same location as that of the removed PC VIC 5 Return and...

Страница 458: ...dure 1 Disconnect the mains AC power cable for the mains power wall outlet 2 Remove the Vivid 4 left and right side covers refer to the Side Covers Removal Procedure on page 8 4 3 Remove the Vivid 4 r...

Страница 459: ...he power supply in position on the BEP as shown in Figure 8 59 Note For the next steps it is recommended to carefully lie the BEP2 on its side to provide easier access 9 Cut any cable ties holding the...

Страница 460: ...cables from the media drives MOD 2 and CD RW 3 13 Remove the screw that fastens the Ground cable and PC VIC in position as shown in Figure 8 62 below 14 Disconnect the power cable from the Hard Disk...

Страница 461: ...BEP cutting any cable ties as necessary 19 Disconnect the mains power cable 8 from the motherboard beside the ECG location 20 Gently slide the power supply further downwards and then out of the BEP ca...

Страница 462: ...oom and arrange each of the cables in the approximate position Carefully feed the cable connectors underneath the data cables 2 Holding the new BEP2 power supply in one hand connect the 4 pin external...

Страница 463: ...ory boards etc are securely in position and that nothing has erroneously become disconnected during the power supply removal procedure 15 Stand the BEP2 upright 16 At the rear of the BEP2 return and l...

Страница 464: ...d 4 ultrasound unit as described in Chapter 3 Installation 8 5 16 3 Hard Disk Removal Procedure 1 Remove the left cover as described Side Covers Removal Procedure section page 8 4 2 Remove the BEP sid...

Страница 465: ...d disk make sure the jumper settings are as shown below 1 Place the new hard disk in position in the hard disk module holder 2 Reconnect all the cables at the rear of the hard disk 3 Fasten each of th...

Страница 466: ...in the BEP2 Side Cover Installation Procedure on page 8 54 7 Return the Vivid 4 left cover as described in the Side Covers Installation Procedure on page 8 4 Note When a new hard disk arrives from the...

Страница 467: ...6 1 3 AC Distribution Box Removal Procedure 1 Remove the Vivid 4 left right and rear covers as described in the Side Covers Removal Procedure on page 8 4 and Rear Cover Removal Procedure on page 8 7 2...

Страница 468: ...n 2 Fasten the AC Distribution Box using the two screws located in the lower bracket of the AC Distribution Box 3 Tighten the screw that connects the Ground cable to the AC Distribution Box 4 Plug in...

Страница 469: ...l Procedure on page 8 7 3 Unplug the internal power cable from the AC Distribution Box 4 Unscrew the Ground cable from the AC Distribution Box 5 Remove the two screws on both the left and the right si...

Страница 470: ...ng all the cables at the back of the unit 4 Open the track by hooking a thumb or finger under the right side of one of the tracks and pulling the side of the track The track clicks open enabling you t...

Страница 471: ...Allen keys as indicated in the cable replacement procedure 8 6 4 1 Preparation Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 8 6 4 2 AC BEP or FE Cable Removal Procedur...

Страница 472: ...ide covers and the rear cover as described in the Side Covers Removal Procedure on page 8 4 and in the Rear Cover Removal Procedure on page 8 7 3 Remove the front end crate as described in the Front E...

Страница 473: ...from the front of the unit and using an M4 Allen key loosen the two cap screws on the gas spring cable retaining bracket and slide the bracket forwards towards you sufficiently to release tension on...

Страница 474: ...cable from the up down swivel handle by carefully sliding the cable end stopper out through the grooved slot on the handle assembly refer to the detail shown in Figure 8 88 on page 8 101 NOTE IMPORTA...

Страница 475: ...sembly carefully aligning the groove on the cable with that of the groove in the slot refer to the detail shown in Figure 8 88 on page 8 101 7 Feed the gas spring cable down into the slot in the cable...

Страница 476: ...16 Return the front end crate as described in the Front End Crate Installation Procedure on page 8 42 17 Return the Vivid 4 side covers and the rear cover as described in the Side Covers Installation...

Страница 477: ...maximum height 2 Remove the control console lower rear cover as described in the Control Console Lower Rear Cover Removal Procedure on page 8 19 3 Remove the control console lower cover as described...

Страница 478: ...cable Then slide the cable end stopper E and cable sleeve F out of the slot in the retaining bracket 9 Working from the front of the unit and using an M4 Allen key loosen the two cap screws on the sw...

Страница 479: ...e cable end stopper out through the grooved slot on the handle assembly refer to the detail shown in Figure 8 88 on page 8 101 NOTE IMPORTANT Before finally removing the swivel release cable pay atten...

Страница 480: ...79 on page 8 96 NOTE For clarification the letters appearing in square brackets in the following step correspond to the letters shown in Figure 8 77 on page 8 95 and Figure 8 78 on page 8 95 6 Workin...

Страница 481: ...ribed in Chapter 3 Installation 8 6 7 3 Up Down Swivel Handle Removal Procedure 1 Remove the control console lower cover rear as described in the Control Console Lower Rear Cover Removal Procedure on...

Страница 482: ...swivel release cable are now fully exposed providing easy access to complete the handle removal procedure 6 Using an M4 Allen key loosen the two cap screws on the cable retaining bracket and slide the...

Страница 483: ...ease the swivel release cable from the up down swivel handle by carefully sliding the cable end stopper out through the grooved slot on the handle assembly refer to the detail shown in Figure 8 88 on...

Страница 484: ...d of the swivel release cable from the vertical axis welding refer to the Swivel Release Cable Removal Procedure on page 8 94 11 Loosen and remove the M6 nut from the handle assembly axis screw see Fi...

Страница 485: ...uring nut against the bracket to hold the cable firmly in position refer also to Figure 8 87 on page 8 100 6 Repeat step 4 to insert the swivel release cable end stopper into the grooved slot on the h...

Страница 486: ...as described in the Front End Crate Removal Procedure on page 8 41 4 Lower the control console to its minimum height NOTE For clarification the letters appearing in square brackets in the following s...

Страница 487: ...and then tighten the screw until the pin L at the lower end of the gas spring is loose enough to be removed and remove the pin NOTE For clarification the letters appearing in square brackets in the f...

Страница 488: ...it the groove of the cable end sleeve C on the hinge F of the gas spring head 6 Turn the cable holder B over the cable end sleeve C and then fasten it with the securing screw A 7 Remove the Hex M8 scr...

Страница 489: ...ved type Part No 2392789 with the same type of wheels refer to the following instructions Front Wheel Removal Procedure 2 on page 8 108 Front Wheel Installation Procedure 2 on page 8 110 8 6 9 1 Tools...

Страница 490: ...s 8 107 Figure 8 94 Releasing the Spring Washer from the Center of the Transverse Arm 3 Remove the four bolts three wrench bolts and one Allen bolt that secure the front wheel to the chassis see Figur...

Страница 491: ...ing the front and the brake in the left locked position preventing movement 1 Remove the Vivid 4 front cover as described in the Front Cover Removal Procedure on page 8 5 2 Remove the locking washer l...

Страница 492: ...el 7 Proceed to step 1 of the Front Wheel Installation Procedure 2 on page 8 110 to install a new front wheel in place of the one removed 8 Repeat step 6 above to remove the other front wheel from the...

Страница 493: ...lways remove and replace one wheel at a time before attempting to remove the second wheel 1 Having removed one wheel from the machine chassis as described in Front Wheel Removal Procedure 2 on page 8...

Страница 494: ...ING When BOTH rear wheels are being removed the machine must be safely supported on two wooden blocks or on a jack capable of supporting the weight of the machine Always remove and replace one wheel a...

Страница 495: ...remove and replace one wheel at a time before attempting to remove the second wheel 1 Having removed one rear wheel position the new wheel on the chassis in the same position as that of the wheel rem...

Страница 496: ...the software CD 3 Hold down the ON OFF button for 5 seconds to shutdown the system The shutdown options are displayed on screen 4 Select Full Shutdown The system powers down 5 Hold down the ON OFF bu...

Страница 497: ...priate flat and Phillips type screw drivers 8 8 1 2 Manpower One person 30 mins 8 8 1 3 Preparations 1 Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 2 Remove the Connect...

Страница 498: ...r on the right side of the Vivid 4 ultrasound unit s top surface below the control console as shown in Figure 8 102 and next to the VCR if present allowing sufficient room to access the cables at the...

Страница 499: ...ribed in the Connector Panels Cover Installation Procedure on page 8 8 7 Reconnect the mains AC power cable from the wall outlet 8 Turn ON the main circuit breaker located on rear of the Vivid 4 9 Tur...

Страница 500: ...r as described in the Connector Panels Cover Removal Procedure on page 8 8 2 Release the B W printer by undoing the fastening belt secured around it and around the VCR if present 3 Disconnect the Remo...

Страница 501: ...The VCR will not operate without initial setup and adjustments as described below Figure 8 105 Mitsubishi Video Cassette Recorder VCR 8 8 2 1 Tools No special tools needed 8 8 2 2 Manpower One person...

Страница 502: ...cable P N 2269430 to the power outlet on the Vivid 4 rear panel and the other end to the AC power socket on the VCR rear panel see Figure 8 107 on page 8 119 3 Using the S video cable P N 2266743 con...

Страница 503: ...ng the power switch on VCR front panel 6 Press Config then from the System Configuration window activate the VCR ECG tab 7 Under the VCR Options list choose Mitsubishi from the VCR model drop down lis...

Страница 504: ...3 VIVID 4 SERVICE MANUAL Chapter 8 Replacement Procedures 8 121 8 From the Video Mode drop down list choose PAL or NTSC according to video system in your region then press OK to apply the changes 9 P...

Страница 505: ...the Enter button to apply your selection To configure the Mitsubishi VCR 1 Press the VCR Play B button on the Vivid 3 control console to display the VCR menus on the Vivid 3 monitor 2 Using the Jog bu...

Страница 506: ...igure 8 113 below then press the Enter button to continue to the next menu 5 Skip Menu3 Figure 8 114 by pressing the Enter button to continue to the next menu 6 Select the RS 232C settings according t...

Страница 507: ...or and Audio switches are in the positions shown in Figure 8 117 below if necessary adjust accordingly then move the Key Lock switch to the ALL position to lock all keys 10 Perform the VCR Test as des...

Страница 508: ...e 8 8 2 If optional shelf P N H45011CN is present refer to the Sony UP 21MD Color Video Printer Removal Procedure on page 8 138 3 Disconnect the RS 232C cable the Audio and S Video IN and OUT cables a...

Страница 509: ...arations 1 Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 2 Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal Procedure on page 8 8 3...

Страница 510: ...d next to the B W Printer if present allowing sufficient room to access the cables at the rear of the VCR Figure 8 121 Video Cassette Recorder VCR Installed in the Vivid 4 Ultrasound Unit Figure 8 120...

Страница 511: ...lled Figure 8 122 Sony VCR Connection Cables AUDIO IN cable From Left Connectors Panel To VCR Audio Input Out R Connector Red Out L Connector Black BLACK CH1 RED CH2 RED BLACK RED BLACK AUDIO OUT cabl...

Страница 512: ...n page 3 45 Note Sony VCR Reset Counter If the Reset Counter button is activated while the VCR is in the Stop Pause or Record state the counter display is reset to zero 0 00 00 8 8 3 5 Sony VCR Remova...

Страница 513: ...s necessary to order the Panasonic VCR RS 232 Adaptor Part No 066E8200 Figure 8 123 Panasonic Video Cassette Recorder VCR 8 8 4 1 Tools No special tools needed 8 8 4 2 Manpower One person 30 mins 8 8...

Страница 514: ...mum height 2 Place the VCR on the Vivid 4 ultrasound unit s top surface left side below the control console as shown in Figure 8 125 below allowing sufficient room to access the cables at the rear of...

Страница 515: ...onic VCR Connection Cables S VIDEO IN cable From Internal IO Module Connector A11 To S Video Input S VIDEO OUT cable From Internal IO Module Connector A10 To S Video Output Power cable From Distributi...

Страница 516: ...ector Panels Cover Installation Procedure on page 8 8 9 Reconnect the mains AC power cable from the wall outlet 10 Turn ON the main circuit breaker located on rear of the Vivid 4 11 Turn ON the Vivid...

Страница 517: ...Release the B W printer by undoing the fastening belt around it together with the VCR if present 3 Disconnect the RS 232C cable with adaptor the Audio and S Video IN and OUT cables and the Power cabl...

Страница 518: ...r on the rear panel Figure 8 128 SONY UP 21MD Color Video Printer 8 8 5 1 Tools Use the appropriate flat and Phillips type screw drivers 8 8 5 2 Manpower One person 30 mins 8 8 5 3 Sony UP 21MD Color...

Страница 519: ...side compartment as shown in Figure 8 129 below 6 Connect the Ground cable hanging out of the Bottom Metal Cover to the Ground screw at the rear of the UP 21MD Color Video Printer as shown in Figure 1...

Страница 520: ...Metal Cover slot see Figure 8 129 on page 8 136 to the UP 21MD Color Video Printer as shown in the diagram in Figure 8 131 below Figure 8 131 Sony UP 21MD Color Video Printer Connection Cables R G B...

Страница 521: ...Color Video Printer Installed on the Vivid 4 8 8 5 4 Sony UP 21MD Color Video Printer Removal Procedure 1 Shut down the Vivid 4 ultrasound unit as described in Chapter 3 Installation 2 Slide the UP 2...

Страница 522: ...0 3 Printer Kit P N 2415435 contains the following items 8 8 6 1 Tools Use the appropriate flat and Phillips type screw drivers 8 8 6 2 Manpower One person 15 mins 8 8 6 3 Preparations 1 Shut down the...

Страница 523: ...age Power IN socket on the rear of the printer as shown in Figure 8 134 below WARNING To avoid electrical hazards the HP 6540 3 USB Deskjet Color Printer must be connected to the AC Supply of the Vivi...

Страница 524: ...click Cancel see Figure 8 135 below Figure 8 135 Found New Hardware Wizard 7 Insert the HP6540 3 driver installation CD P N 2415381 provided with the printer into the CD ROM drive 8 Browse the CD Dri...

Страница 525: ...8 Peripherals 9 When the Add Printer Wizard opens click Next see Figure 8 138 10 When prompted verify that Local Printer is selected If not activate the Local printer radio button as shown in Figure...

Страница 526: ...VICE MANUAL Chapter 8 Replacement Procedures 8 143 11 Click Next 12 From the displayed list of available ports choose one of the available USB ports as shown in Figure 8 140 13 Click Next Figure 8 139...

Страница 527: ...pherals 14 From the list of Printers highlight the HP Deskjet 6500 Series see Figure 8 141 15 Click Next 16 When prompted choose Yes if you want this printer to be the default printer see Figure 8 142...

Страница 528: ...es 8 145 18 When the following message appears verify that Do not share this printer is selected as shown in Figure 8 143 19 Click Next 20 In response to the Print a test page message choose No and th...

Страница 529: ...rals The Add Printer Wizard displays the following message 21 Click Finish to close the Wizard The Printer Properties Dialog opens with the General Tab activated as shown in Figure 8 146 22 Click on P...

Страница 530: ...message is displayed 23 Confirm that test page was printed properly then click OK The Printer Properties Dialog opens with the Ports Tab activated as shown in Figure 8 148 24 Confirm that a USB port...

Страница 531: ...ripherals The Printer Properties Dialog opens with the Advanced Tab activated as shown in Figure 8 149 25 Click on Printing Defaults The Printing Defaults dialog box opens with the Printing Shortcuts...

Страница 532: ...etc to your liking then click OK The following message is displayed 27 Press any key to reboot the system and then eject the CD from the CD tray 28 Press the Config button on control console and from...

Страница 533: ...tly Installed Printers list then click OK 8 8 6 5 HP 6540 3 USB Deskjet Color Printer Removal Procedure 1 Turn OFF power to the USB printer 2 Unplug the AC power cable from the Power IN socket at the...

Страница 534: ...n 9 1 Overview 9 1 1 Purpose of Chapter 9 This chapter gives you an overview of replacement parts for the Vivid 4 ultrasound unit Table 9 1 Contents in Chapter 9 Section Description Page Number 9 1 Ov...

Страница 535: ...st of Abbreviations Assy Assembly BEP Back End Processor Ctrl Control FEP Front End Processor FRU 1 Replacement part available in parts hub FRU 2 Replacement part available from the manufacturer lead...

Страница 536: ...1 Mechanical Hardware Parts Figure 9 1 Mechanical Hardware Parts Diagram 1 Table 9 2 Mechanical Hardware Parts Diagram 1 Item Part Name Part Number Description Qty FRU 101 Cover Rear 2258594 1 2 102 C...

Страница 537: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL 9 4 Section 9 3 Renewal Parts Lists and Diagrams Figure 9 2 Mechanical Hardware Parts Diagram 2 201 202 203 204 205 206 207...

Страница 538: ...Part Name Part Number Description Qty FRU 201 Cover Right 2258600 1 1 202 Cover Top Lower Section 2258602 1 2 203 Control Console Upper Cover Front V4 2405642 1 1 204 Control Console Upper Cover Rear...

Страница 539: ...3 Table 9 4 Mechanical Hardware Parts Diagram 3 Item Part Name Part Number Description Qty FRU 301 Filter Air 2301690 1 2 302 Cover Gas Spring Sleeve V4 2261575 2 1 2 303 Probe Cable Holder V4 226384...

Страница 540: ...Diagram 4 Table 9 5 Mechanical Hardware Parts Diagram 4 Item Part Name Part Number Description Qty FRU 501 Wheel Castor Rear 2389102 Rear 2 1 502 Wheel Castor Front 2392789 Front with brake and trans...

Страница 541: ...Illustrated Item Part Name Part Number Description Qty FRU 601 Swivel Release Cable 2392103 1 2 602 Monitor Service Cover 2392137 1 2 603 Screw Kit 2253052 1 1 604 Keycaps Kit V4 2394000 2 1 1 605 Ve...

Страница 542: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 9 Renewal Parts 9 9 9 3 2 AC System Parts Figure 9 5 AC System Parts 701 702 703 704 705 706 707 708...

Страница 543: ...reaker 2252994 1 1 702 AC Voltage Distribution Box 2399304 2 1 1 703 Connector AC Voltage Primary 100 120V 2253054 100 120V 1 2 704 Connector AC Voltage Primary 200 240V 2253056 200 240V 1 2 705 Conne...

Страница 544: ...Front End Parts Diagram 1 Table 9 8 Front End Parts Diagram 1 Item Part Name Part Number Description Qty FRU 801 Front End Crate 2253030 2 including back plan MB 1 2 802 Front End Crate 2253030 3 syst...

Страница 545: ...d Parts Diagram 2 Item Part Name Part Number Description Qty FRU 1001 PCB Front Board and TR4 Assembly 2253033 3 FB 3Con V7 1 1 1002 PCB TR4 Board 2418129 16 1 1003 PCB Multiplexer MUX 2253038 3 1 1 1...

Страница 546: ...3 Item Part Name Part Number Description Qty FRU 1011 PCB RF Tissue RFT MLA2 FB200992 Applies to systems with RFT configuration only 1 1 1012 PCB Beamformer BF 64ch FB200900 1 1 1013 PCB Image Port 3...

Страница 547: ...and Diagrams 9 3 4 Back End Parts Figure 9 9 Back End Parts Diagram 1 Table 9 11 Back End Parts Diagram 1 Item Part Name Part Number Description Qty FRU 1021 Keyboard Trackball 2274623 1 1 1022 Keybo...

Страница 548: ...bly Spanish 2393953 4 1 2 Keyboard Overlay Assembly Portuguese 2393953 5 1 2 Keyboard Overlay Assembly Italian 2393953 6 1 2 Keyboard Overlay Assembly Russian 2393953 8 1 2 Keyboard Overlay Assembly D...

Страница 549: ...attery 2300858 1 1 1052 PCB Plug Scan Card 2316279 1 1 1053 MO Drive 5 25 2277189 This hardware is an option that should be ordered P N H45011CF 1 1 1054 Footswitch 066E0007 1 2 1055 BEP Vivid 4 Assy...

Страница 550: ...GE MEDICAL SYSTEMS DIRECTION 2392751 100 REVISION 3 VIVID 4 SERVICE MANUAL Chapter 9 Renewal Parts 9 17 Figure 9 12 Back End Parts Diagram 4 1061 1062 1063 1064 1065 1066 1067 1068 1069...

Страница 551: ...PC VIC 1 1 1062 Hard Disk Drive HDD 40 to 100 GB 2408479 1 1 1063 Floppy Drive 2346034 1 2 1064 CD Drive RW 2413449 1 1 1065 PCB SCSI Card 2271149 2 This hardware is an option that should be ordered...

Страница 552: ...Cable VGA VIC PC 2252965 1 2 1103 Cable Input Box AC Distribution Box 2252966 1 2 1104 Cable Soft Shutdown 2252967 1 2 1105 Cable Ground VIC 2252968 1 2 1106 Cable Ground Back End 2252969 1 2 1107 Cab...

Страница 553: ...k End Matrix Panel 2252978 1 2 1115 Cable ECG Back End Internal 2252979 1 2 1116 Cable Patient I O int ECG ECG Phono 2252980 1 2 1117 Cable ECG Int Footswitch Keyboard Control 2252982 1 2 1118 Cable C...

Страница 554: ...C Box 2254089 1 2 1128 Cable ECG BEP Cable USA External 2256477 1 1129 Cable ECG BEP Cable Europe External 2256478 1 1 1130 Cable External ECG with Echo Stress Option 2256685 1 2 1131 Cable Internal A...

Страница 555: ...S232 Adaptor for Panasonic VCR 066E8200 1 1 1138 Cable Video IN Rear Panel 2266743 1 2 1139 Cable Video OUT Rear Panel 2266744 1 2 1140 Cable Audio OUT Rear Panel 2266745 1 2 1141 Cable Audio IN Rear...

Страница 556: ...tion Box 2292972 1 N 1150 Cable USB Cable 2297905 1 2 1151 Cable ECG Internal COM 2 2298089 1 N 1152 Cable Power Peripherals and Printer 2300857 1 2 1153 Cable Cable Power Asia 2269428 2 1 1 1154 Cabl...

Страница 557: ...EVISION 3 VIVID 4 SERVICE MANUAL 9 24 Section 9 3 Renewal Parts Lists and Diagrams 9 3 6 Software Table 9 16 Software Item Part Name Part Number Description Qty FRU 1301 S W Kit contains s w for syste...

Страница 558: ...08 Probe 5S Sector H40422LA H4901RA 1 1 1409 Probe 10L 739L Linear H40412LG 1 1 1410 Probe 39A 12L Linear H40412LH 1 1 1411 Biopsy Kit for 10L 12L E83885MM 1 N 1412 Probe 10S H4901PC 1 1 1413 Probe P5...

Страница 559: ...MD3000E S VHS RS232C PAL H45011PT 1 2 1502 VCR Mitsubishi HS MD3000U S VHS RS232C NTSC H45011PN 1 2 1503 VCR Cassette S VHS H45001PG 1 N 1504 Color Printer Tray V4 2393556 1 2 1505 Sony UP 21MD PAL N...

Страница 560: ...77 2252975 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2269429 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2...

Страница 561: ...16 2252977 2252975 2252982 2258801 2252964 2253011 2252976 2252981 2253009 2253015 2292972 2276874 2300847 2253014 2276875 2253080 2253079 RS232 1 2266742 2269430 C B L F AC INPUT BOX 2252994 On off 2...

Страница 562: ...n the backplane the AC distribution box and the DC distribution box that could be dangerous Be sure to disconnect the system power plug and to open the main circuit breaker before you remove any parts...

Страница 563: ...canner or nearby 10 2 2 Quality Assurance In order to gain accreditation from organizations such as the American College of Radiology USA it is the customer s responsibility to have a Quality Assuranc...

Страница 564: ...nner as recommended in the Periodic Maintenance Schedule in order to retain the high levels of safety dependability and system performance Your GE Medical Systems Service Representative has an in dept...

Страница 565: ...rective maintenance or as required by your facility s QA program Peripheral Current Leakage Checks Also after corrective maintenance or as required by your facility s QA program Surface Probe Current...

Страница 566: ...abrasions Equipment for loose or missing hardware The keyboard for defects Brakes for defective operation DANGER To avoid electrical shock hazard do not remove the panels or covers from the machine T...

Страница 567: ...le wrist strap 3M 3051 conductive Ground cord Anti Static Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Air Filter 2256176 Air intake Safety Analyzer 46 285652G1 DALE 600 KIT or equivalent for...

Страница 568: ...if they have any problems or questions about the equipment 2 Paperwork Fill in the appropriate details at the top of the Vivid 4 Preventative Maintenance Inspection Certificate see page 10 31 Note al...

Страница 569: ...Mode operation Check the basic system controls that affect this mode of operation 5 Applicable Software Optionsa a Some software may be considered as standard depending upon system model configuratio...

Страница 570: ...ssary 3 Color Printer Verify hardcopy output of the Color video page printer Clean heads and covers if necessary 4 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 5 InS...

Страница 571: ...lps to maintain a clean working environment Prior to cleaning turn OFF the power to the system Use a fluid detergent in warm water on a soft damp cloth to carefully wipe the entire system Be careful n...

Страница 572: ...the air filter replacement part number Table 10 9 Air Filter Cleaning Step Item Description 1 Air Filter Remove the air filter located at the front of the scanner above the probe connector sockets Cl...

Страница 573: ...MOD disks must be stored away from dust and cigarette smoke Do not use alcohol or benzene to clean the MOD cartridge 10 CD Drive Clean the drive head and media with the vendor supplied cleaning kit A...

Страница 574: ...0 13 10 5 6 Diagnostic Checks Optional To complete the periodic maintenance checks perform the diagnostic tests as described in Chapter 7 Diagnostics Troubleshooting Review the system error log for an...

Страница 575: ...heck of the probe pins and system sockets before plugging in a probe Avoid pulling pinching or kinking the probe cable since a damaged cable may compromise the electrical safety of the probe To avoid...

Страница 576: ...TEE and Interoperative probes often have special usage considerations always refer to the individual probe manufacturers handling instructions user manual 10 6 4 Probe Cleaning 10 6 4 1 Basic Probe Cl...

Страница 577: ...ogen free probe sheaths Probes used during neurological surgery must NOT be sterilized with liquid chemical sterilants because of the possibility of neurotoxic residue remaining on the probe 10 6 4 2...

Страница 578: ...will receive and or open the package NOTE The US Department of Transportation DOT has ruled that items that were saturated and or dripping with human blood that are now caked with dried blood or whic...

Страница 579: ...under test must not be connected to other electrical equipment Remove all interconnecting cables and wires The operator and patient must not come into contact with the unit while the tests are being...

Страница 580: ...rmal Condition Open Ground Reverse Polarity Open Neutral Mains Applieda a Mains Applied is the Sink Leakage Test In this test mains supply voltage is supplied to the part or equipment to determine the...

Страница 581: ...the Grounding protective earth conductor have been reversed If later tests indicate high current leakages this should be suspected as a possible cause and the outlet wiring should be visually inspecte...

Страница 582: ...hese steps to test the Ground wire resistance 1 Turn the Vivid 4 unit OFF 2 Connect one of the Ohmmeter probes to the tested AC wall outlet Ground pin 3 Using the other Ohmmeter probe touch the expose...

Страница 583: ...powered ON Be sure to turn the unit power OFF before switching them using the Polarity switch and or the Neutral switch 10 7 5 2 Generic Procedure The test verifies the isolation of the power line fro...

Страница 584: ...le 10 14 Record all data on the Vivid 4 Preventative Maintenance Inspection Certificate see page 10 31 Table 10 17 Typical Data Sheet for Chassis Source Current Leakage Test Unit Power Tester Polarity...

Страница 585: ...s should be made with the Ground Open and Closed with the Vivid 4 unit ON and OFF and with the power line polarity Normal and Reversed Record the highest reading For each combination the operating con...

Страница 586: ...d Patient Lead Sink Leakage Isolation Test NOTE Refer to the procedure in the IEC 60601 1 CAUTION Line voltage is applied to the ECG leads during this test To avoid the risk of possible electric shock...

Страница 587: ...Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Current Leakage Test and Patient Lead to Lead Current Leakage Test 115V 10uA 10uA 220 240V 500uA 10uA Table 10 19 Maxim...

Страница 588: ...made with the Ground OPEN and CLOSED with power line polarity Normal and Reversed and with the unit OFF and ON For each combination the probe must be active to find the worst case condition NOTE Each...

Страница 589: ...be into the meter EXTERNAL connector 4 Set the meter s FUNCTION switch to EXTERNAL position 5 Connect the probe for test with the connector of the console 6 Add the saline probe and the imaging area o...

Страница 590: ...10 31 CAUTION When power to the unit is ON never switch the Polarity and the status of Neutral Be sure to turn OFF power to the unit before switching them using the POLARITY switch and or the NEUTRAL...

Страница 591: ...lies with the probe or the scanner NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe current leakages are normal from probe to probe Other variations...

Страница 592: ...ormat Model Number Serial Number Probe 6 Frequency Scan Format Model Number Serial Number Probe 7 Frequency Scan Format Model Number Serial Number Probe 8 Frequency Scan Format Model Number Serial Num...

Страница 593: ...Source Current Leakage CRT Patient Lead Source Leakage Lead to Ground Patient Lead Source Leakage Lead to Lead Patient Lead Source Leakage Isolation Peripheral 1 Current Leakage Peripheral 1 Ground C...

Страница 594: ...mericas Fax 414 544 3384 P O Box 414 Milwaukee Wisconsin 53201 0414 U S A GE Medical Systems Europe Fax 33 1 40 93 33 33 Paris France GE Medical Systems Asia Fax 65 291 7006 Singapore GE Vingmed Ultra...

Отзывы: