Chapter 3. Making Measurements
64
USM Go Operator’s Manual
3.1.1
Positioning the Gates
Use the procedures in this section to set the vertical and horizontal position of Gate A and Gate B. Remember that gate position has the
following effects on instrument performance:
•
A-Scan echoes on the right side of the display screen represent features that occur at a greater depth from the test-material surface than
those on the left side of the display screen. Therefore, moving a gate to the right means that the gate is evaluating a deeper portion of the
test material.
•
A wider gate setting means that the gate is evaluating a thicker portion of the test material.
•
Increasing the vertical height, called the threshold, of a gate means that only reflected signals of sufficiently large amplitude will cross
the gate.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......