![GE Ultrasonic USM Go Скачать руководство пользователя страница 119](http://html.mh-extra.com/html/ge/ultrasonic-usm-go/ultrasonic-usm-go_operators-manual_128490119.webp)
Chapter 3. Making Measurements
USM Go Operator’s Manual
101
3.13.5 Locks and Error Messages
As long as a valid reference echo is stored, no functions can be changed which could cause an incorrect
DGS
evaluation. If an attempt is
made to change such a function, the following error message appears:
•
“Function Locked - DGS Reference has been recorded”
Likewise, the
DGS
evaluation must be switched
OFF
and the reference echo deleted when selecting a new probe (e.g. for a new test
application).
3.13.6 Validity of the DGS Method
Echo amplitude evaluations according to the
DGS
method are only reliable and reproducible in cases when:
•
The reference echo is received from the test object if possible. If this is not possible, it should be verified that the reference block is made
of the same material as the test object.
•
The evaluation is carried out using the same probe which was also used for recording the reference echo. Another probe of the same type
can be used only after recording a new reference echo.
•
Echo amplitudes for reflector distances smaller than half of the probe near-field length are subject to extreme variation, for physical
reasons due to interference phenomena effecting the area. Thus, evaluation results may fluctuate more than the usually permissible
±2 dB. An evaluation according to the
DGS
method is possible, but it is not recommended in such a case.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......