![GE Ultrasonic USM Go Скачать руководство пользователя страница 123](http://html.mh-extra.com/html/ge/ultrasonic-usm-go/ultrasonic-usm-go_operators-manual_128490123.webp)
Chapter 3. Making Measurements
USM Go Operator’s Manual
105
3.15 AWS D1.1 Weld Rating Evaluation Mode (cont.)
Before activating the
AWS D1.1/D1.5
weld rating mode, be sure that all instrument settings are properly adjusted for the specific
measurement application. Then access the
AWS D1.1
submenu in the
ACQUIRE
menu, and follow this procedure:
1.
Apply couplant and couple the probe to a suitable reference test standard.
2.
Ensure that Gate A is positioned over the desired echo. Then, adjust the gain until the peak of the desired reference echo reaches the
desired amplitude. Note that if the echo peak amplitude (A%A) does not fall between 10% and 90% of FSH, the point will not be
accepted.
3.
Use the
B REFERENCE
function to define the reference dB level.
4.
To evaluate a reflector in a test piece, couple the probe to the test piece. Now adjust the Gate A position, if required, so that it is over the
desired echo.
5.
Adjust the gain until the peak of the test-piece echo reaches the desired screen height.
Note:
The dB value recorded for A-Indication is automatically adjusted based on the amplitude difference between it and the value
recorded for
B-REF
. It will usually not match the instrument dB gain setting.
6.
With the
AWS D1.1 SETUP
submenu displayed, note that the
A
,
C
, and
D
parameters automatically update to match the Gate A
triggering echo. Use the
A INDICATION
function to
HOLD
the current parameters. Note that the displayed
RESULTS
continue to
update while the
A
,
C
, and
D
parameters are held.
Содержание Ultrasonic USM Go
Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259801 Rev 1 July 2009 USM Go Operator s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page proceed to next page ...
Страница 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 80: ...Chapter 2 Instrument Setup 62 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 130: ...Chapter 3 Making Measurements 112 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 152: ...Chapter 5 Data Recorder Files 134 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Страница 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Страница 189: ......